https://arxiv.org/api/lqYc2dGcK6nCijqPbcAHDIJAdTc 2026-06-23T00:49:28Z 8985 615 15 http://arxiv.org/abs/2507.20413v2 Electromagnetic Noise Characterization and Suppression in Low-Field MRI Systems 2025-12-12T17:32:19Z Purpose: Low-field MRI systems operate at single MHz-range frequencies, where signal losses are primarily dominated by thermal noise from the radio-frequency (RF) receive coils. Achieving operation close to this limit is essential for maximizing imaging performance and signal-to-noise ratio (SNR). However, electromagnetic interference (EMI) from cabling, electronics, and patient loading often degrades system performance. Our goal is to develop and validate a practical protocol that guides users in identifying and suppressing electromagnetic noise in low-field MRI systems, enabling operation near the thermal noise limit. Methods: We present a systematic, stepwise methodology that includes diagnostic measurements, hardware isolation strategies, and good practices for cabling and shielding. Each step is validated with corresponding noise measurements under increasingly complex system configurations, both unloaded and with a human subject present. Results: Noise levels were monitored through the incremental assembly of a low-field MRI system, revealing key sources of EMI and quantifying their impact. Final configurations achieved noise within 1.5x the theoretical thermal bound with a subject in the scanner. Image reconstructions illustrate the direct relationship between system noise and image quality. Conclusion: The proposed protocol enables low-field MRI systems to operate close to fundamental noise limits in realistic conditions. The framework also provides actionable guidance for the integration of additional system components, such as gradient drivers and automatic tuning networks, without compromising SNR. 2025-07-27T21:13:27Z This is a protocol for optimizing the performance of low-field MRI scanners. If you use it, please give us feedback! Magn Reson Med. 2026; 95(5): 3000-3007 Teresa Guallart-Naval José M. Algarín Joseba Alonso 10.1002/mrm.70235 http://arxiv.org/abs/2512.11467v1 Mastering Preclinical Fast MRSI: From Setup to Execution 2025-12-12T11:08:36Z MR experiments are essential for studying brain metabolism, yet preclinical 1H-MRSI remains underdeveloped, with significant limitations in SNR, acquisition speed, and automated data processing. Although recent advances-such as accelerated sequences, denoising strategies, and ultra-high-field systems-have begun to reduce these barriers, preclinical MRSI still lags far behind the human research field in accessibility and routine use. Based on our expertise, we have created this guide that outlines a complete workflow for acquiring and analyzing high-quality fast MRSI data in rodent brains at ultra-high fields (9.4T and 14.1T), enabling novice users to perform reliable experiments using optimized MRSI sequences (FID-MRSI, SE-MRSI, and PRESS-MRSI) and standardized processing pipelines, while also highlighting strategies to further improve acquisition speed, coverage, and reproducibility. Overall, this paper provides a strong foundation for future methodological advances that will expand metabolic imaging capabilities and deepen insights into brain function and disease. 2025-12-12T11:08:36Z Gianna Nossa Eloïse Mougel Brayan Alves Tan Toi Phan Alessio Siviglia Thi Ngoc Anh Dinh Thanh Phong Lê Bernard Lanz Cristina Cudalbu http://arxiv.org/abs/2503.16741v5 CTorch: PyTorch-Compatible GPU-Accelerated Auto-Differentiable Projector Toolbox for Computed Tomography 2025-12-11T23:40:15Z This work introduces CTorch, a PyTorch-compatible, GPU-accelerated, and auto-differentiable projector toolbox designed to handle various CT geometries with configurable projector algorithms. CTorch provides flexible scanner geometry definition, supporting 2D fan-beam, 3D circular cone-beam, and 3D non-circular cone-beam geometries. Each geometry allows view-specific definitions to accommodate variations during scanning. Both flat- and curved-detector models may be specified to accommodate various clinical devices. CTorch implements four projector algorithms: voxel-driven, ray-driven, distance-driven (DD), and separable footprint (SF), allowing users to balance accuracy and computational efficiency based on their needs. All the projectors are primarily built using CUDA C for GPU acceleration, then compiled as Python-callable functions, and wrapped as PyTorch network module. This design allows direct use of PyTorch tensors, enabling seamless integration into PyTorch's auto-differentiation framework. These features make CTorch an flexible and efficient tool for CT imaging research, with potential applications in accurate CT simulations, efficient iterative reconstruction, and advanced deep-learning-based CT reconstruction. 2025-03-20T23:06:51Z Xiao Jiang Grace J. Gang J. Webster Stayman http://arxiv.org/abs/2509.13644v3 Sensitivity of literature T1 mapping methods to the underlying magnetization transfer parameters 2025-12-11T22:17:52Z Purpose: Magnetization transfer (MT) has been identified as the principal source of $T_1$ variability in the MRI literature. This study assesses the sensitivity of established $T_1$ mapping techniques to variations in the underlying MT parameters. Methods: For each $T_1$-mapping method, the observed $T_1$ was simulated as a function of the underlying MT parameters $p_i^\text{MT}$, corresponding to different brain regions of interest (ROIs) at 3T. As measures of sensitivity, the derivatives $\partial T_1^\text{observed} / \partial p_i^\text{MT}$ were computed and analyzed with a linear mixed-effects model as a function of $p_i^\text{MT}$, ROI, pulse sequence type (e.g., inversion recovery, variable flip angle), and the individual sequences. Results: The analyzed $T_1$-mapping sequences have a considerable sensitivity to changes in the semi-solid spin pool size $m_0^\text{s}$, $T_1^\text{f}$ of the free, $T_1^\text{s}$ of the semi-solid spin pool, and the (inverse) exchange rate $T_\text{x}$. All derivatives vary considerably with the underlying MT parameters and between pulse sequences. In general, the derivatives cannot be determined by the sequence type, but rather depend on the implementation details of the sequence. One notable exception is that variable-flip-angle methods are, in general, more sensitive to the exchange rate than inversion-recovery methods. Conclusion: Variations in the observed $T_1$ can be caused by several underlying MT parameters, and the sensitivity to each parameter depends on both the underlying MT parameters and the sequence. 2025-09-17T02:41:30Z Jakob Assländer http://arxiv.org/abs/2512.11017v1 Integrating Uncertainty Quantification into Computational Fluid Dynamics Models of Coronary Arteries Under Steady Flow 2025-12-11T18:08:10Z Computational models are continuously integrated in the clinical space, where they support clinicians in disease diagnosis, prognosis, and prevention strategies. While assisting in clinical space, these computational models frequently use deterministic approaches, where the inherent (aleatoric) variability of input parameters is ignored. This questions the credibility and often hinders the clinical adoption of these computational models. Therefore, in this study, we introduced uncertainty quantification in the computational fluid dynamics models of the left main coronary artery to analyze the influence of input hemodynamics parameters on wall shear stress (WSS). UncertainSCI was used, where an emulator was built using polynomial chaos expansion between the input parameters and the output quantity of interest, and the output sensitivities and statistics were directly extracted from the emulator. The uncertainty-informed framework was first applied to an analytical solution of the Navier-Stokes equation (Poiseuille flow) and then to a patient-specific model of the left main coronary artery. Different input hemodynamics parameters are considered, such as pressure, viscosity, density, velocity, and radius, whereas wall shear stress was considered as our output quantity of interest. The results suggest that velocity dominated the variability in WSS in the analytical model (~79%), whereas viscosity dominated in the patient-specific model (~59%). The results further suggest that out of all the Sobol indices interactions, unary interactions were the most dominant ones, contributing ~93.2% and ~99% for the analytical and patient-specific model, respectively. This study will enhance confidence in computational models, facilitating their adoption in the clinical space to improve decision-making for coronary artery disease diagnosis, prognosis, and therapeutic strategies. 2025-12-11T18:08:10Z Muhammad Usman Peter N. Castillo Akil Narayan Lucas H. Timmins 10.1115/1.4071773 http://arxiv.org/abs/2512.10745v1 PMB-NN: Physiology-Centred Hybrid AI for Personalized Hemodynamic Monitoring from Photoplethysmography 2025-12-11T15:32:50Z Continuous monitoring of blood pressure (BP) and hemodynamic parameters such as peripheral resistance (R) and arterial compliance (C) are critical for early vascular dysfunction detection. While photoplethysmography (PPG) wearables has gained popularity, existing data-driven methods for BP estimation lack interpretability. We advanced our previously proposed physiology-centered hybrid AI method-Physiological Model-Based Neural Network (PMB-NN)-in blood pressure estimation, that unifies deep learning with a 2-element Windkessel based model parameterized by R and C acting as physics constraints. The PMB-NN model was trained in a subject-specific manner using PPG-derived timing features, while demographic information was used to infer an intermediate variable: cardiac output. We validated the model on 10 healthy adults performing static and cycling activities across two days for model's day-to-day robustness, benchmarked against deep learning (DL) models (FCNN, CNN-LSTM, Transformer) and standalone Windkessel based physiological model (PM). Validation was conducted on three perspectives: accuracy, interpretability and plausibility. PMB-NN achieved systolic BP accuracy (MAE: 7.2 mmHg) comparable to DL benchmarks, diastolic performance (MAE: 3.9 mmHg) lower than DL models. However, PMB-NN exhibited higher physiological plausibility than both DL baselines and PM, suggesting that the hybrid architecture unifies and enhances the respective merits of physiological principles and data-driven techniques. Beyond BP, PMB-NN identified R (ME: 0.15 mmHg$\cdot$s/ml) and C (ME: -0.35 ml/mmHg) during training with accuracy similar to PM, demonstrating that the embedded physiological constraints confer interpretability to the hybrid AI framework. These results position PMB-NN as a balanced, physiologically grounded alternative to purely data-driven approaches for daily hemodynamic monitoring. 2025-12-11T15:32:50Z Yaowen Zhang Libera Fresiello Peter H. Veltink Dirk W. Donker Ying Wang 10.1016/j.cmpb.2026.109479 http://arxiv.org/abs/2511.14500v2 Composition-Dependent Properties of $\mathrm{Ce_{x}La_{0.95-x}Tb_{0.05}F_{3}}$ Nanopowders Tailored for X-Ray Photodynamic Therapy and Cathodoluminescence Imaging 2025-12-11T11:47:27Z This study investigates the synthesis and luminescence behavior of $\mathrm{Ce_{x}La_{0.95-x}Tb_{0.05}F_{3}}$ nanoparticles with varying $\mathrm{Ce^{3+}}$ content. The materials were prepared via a wet chemical route and thermally annealed to improve crystallinity and reduce defects. Phase composition and structural parameters were examined by X-ray diffraction (XRD), while elemental composition was determined by X-ray fluorescence (XRF). Cathodoluminescence (CL) intensity mapping was used to evaluate emission uniformity and monitor the degradation of luminescence under electron beam exposure. Photoluminescence (PL) and radioluminescence (RL) spectroscopy confirmed energy transfer from $\mathrm{Ce^{3+}}$ to $\mathrm{Tb^{3+}}$ ions. Luminescence intensities were found to depend strongly on both Ce content and thermal treatment. The results contribute to the understanding of defect-related quenching mechanisms and are relevant for the design of rare-earth-based luminescent nanomaterials for biomedical applications. 2025-11-18T13:46:32Z Xenie Lytvynenko Marie Urbanová Ondřej Lalinský Vilém Vojta Jan Bárta Lenka Prouzová Procházková Václav Čuba 10.1016/j.radmeas.2025.107536 http://arxiv.org/abs/2512.10555v1 Event-sparse stack denoising for 4D-STEM applications 2025-12-11T11:39:12Z We introduce a denoising method for four-dimensional scanning transmission electron microscopy (4D-STEM) that relies on processing local, scan position-independent electron event-sparse data stacks, called event-sparse stack denoising. This method adds an extra time dimension during data collection by recording multiple electron event-sparse diffraction patterns. The resulting datasets are effectively five-dimensional, referred to as locally time-resolved STEM (LTR-STEM). Diffraction data stacks at each scan position are processed using one of two sparsity denoising pipelines: 1) the density-based spatial clustering of applications with noise (DBSCAN) algorithm followed by multi-step persistence thresholding, or 2) sparse principal component analysis (sparse PCA), followed by single-step thresholding. Both methods perform well for diffraction data denoising, as shown by simulated peak signal-to-noise ratio (PSNR) curves, denoised experimental data for virtual imaging, and application-specific denoising for defect detection. PSNR analysis indicates that sparsity-denoised 4D-STEM data reaches the same PSNR as raw data at approximately 16% of the exposure time, demonstrating comparable image quality with a lower dose. In defect detection, a 4.1x increase in sensitivity to relative radial disk shift is observed in the denoised data. Moreover, the LTR-STEM technique may be used to inspect material degradation by tracking changes in diffraction disk intensity, allowing for critical dose estimation and exposure-selective imaging. 2025-12-11T11:39:12Z Gregory Nordahl Rebekka Klemmt Espen Drath Bøjesen http://arxiv.org/abs/2512.10377v1 KRAS G12D protein screening for pancreatic cancer clinical trials using an AlGaN/GaN high electron mobility transistor biosensor 2025-12-11T07:41:42Z Clinical trials screening KRAS G12D protein for 30 pancreatic ductal adenocarcinoma (PDAC) patients and 30 healthy donors were conducted utilizing an AlGaN/GaN high electron mobility transistor (HEMT) biosensor. All resistance change ratios of PDAC patients are higher than the standard deviation above the mean resistance change ratio obtained from all healthy donors. The results demonstrate the effectiveness of the HEMT biosensor and reveal its potential for early detection of pancreatic cancer with KRAS G12D protein screening. 2025-12-11T07:41:42Z Sheng-Ting Hung Department of Physics, National Sun Yat-sen University, Kaohsiung, Taiwan Cheng Yan Lee Department of Physics, National Sun Yat-sen University, Kaohsiung, Taiwan Chen-Yu Lien Department of Physics, National Sun Yat-sen University, Kaohsiung, Taiwan Cheng-Hsuan Chan Department of Physics, National Sun Yat-sen University, Kaohsiung, Taiwan Ya-Han Yang Department of Surgery, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan Quark Yungsung Chen Industry Academia Innovation School, National Yang Ming Chiao Tung University, Hsinchu, Taiwan Department of Physics and Texas Center for Superconductivity, University of Houston, Houston, TX, USA Kuang-Hung Cheng Institute of Biomedical Sciences, National Sun Yat-sen University, Kaohsiung, Taiwan Kung-Kai Kuo Department of Surgery, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan Department of Surgery, E-DA Healthcare Group E-DA Dachang Hospital, Kaohsiung, Taiwan Li-Wei Tu Department of Physics, National Sun Yat-sen University, Kaohsiung, Taiwan Ching-Wen Chang Institute and Undergraduate Program of Electro-Optical Engineering, National Taiwan Normal University, Taipei, Taiwan 10.1109/JSEN.2026.3669133 http://arxiv.org/abs/2512.09173v1 The choice of viscous or viscoelastic models affects attenuation and velocity determination in simplified skull-mimicking digital phantoms 2025-12-09T22:38:10Z Simulation-guided transcranial focused ultrasound therapies rely on estimating skull acoustic properties from pretreatment imaging. Typical clinical resolution (0.5 mm isotropic) cannot resolve bone microstructure, making the acoustic properties underdetermined and sensitive to modeling assumptions. Here, we examine how viscous and viscoelastic models predict changes in attenuation and phase velocity due to microstructure. Using viscous and viscoelastic k-Wave implementations, we simulated transmission of a broadband 625 kHz tone burst (250 kHz-1 MHz) through skull-mimicking digital phantoms. The phantoms contained spherical pores (0.1-1.0 mm diameter) randomly embedded within cortical bone (2.5%-90% porosity). Virtual sensors measured attenuation and phase velocity using a time-distance matrix approach. Both models predict increased attenuation with increasing pore size at a fixed porosity, but differ in the strength and porosity dependence of this relationship. The viscoelastic model generally predicts attenuation peaks at higher porosities than the viscous model. For 1.0 mm pores, the viscous peak (1.98 Np/cm) occurs at 20% porosity, while the viscoelastic peak (2.98 Np/cm) occurs at 70%. Phase velocity decreases with pore size for both models, though the viscoelastic predictions are less sensitive to pore size. These results demonstrate that viscous and viscoelastic models exhibit distinct attenuation and phase-velocity behavior for idealized bone microstructures. While both indicate that microstructure has a strong impact on attenuation, it has a lesser effect on phase velocity for the viscoelastic model compared to the viscous model. This work highlights the importance of acoustic model choice when estimating skull acoustic properties from computed tomography images. Future work will identify which acoustic model best represents ultrasound propagation through skull microstructure. 2025-12-09T22:38:10Z Submitted to Physics in Medicine and Biology Samuel Clinard Taylor Webb Henrik Odéen Dennis L. Parker Douglas A. Christensen http://arxiv.org/abs/2507.15078v2 PET Image Reconstruction Using Deep Diffusion Image Prior 2025-12-09T15:49:40Z Diffusion models have shown great promise in medical image denoising and reconstruction, but their application to Positron Emission Tomography (PET) imaging remains limited by tracer-specific contrast variability and high computational demands. In this work, we proposed an anatomical prior-guided PET image reconstruction method based on diffusion models, inspired by the deep diffusion image prior (DDIP) framework. The proposed method alternated between diffusion sampling and model fine-tuning guided by the PET sinogram, enabling the reconstruction of high-quality images from various PET tracers using a score function pretrained on a dataset of another tracer. To improve computational efficiency, the half-quadratic splitting (HQS) algorithm was adopted to decouple network optimization from iterative PET reconstruction. The proposed method was evaluated using one simulation and two clinical datasets. For the simulation study, a model pretrained on [$^{18}$F]FDG data was tested on [$^{18}$F]FDG data and amyloid-negative PET data to assess out-of-distribution (OOD) performance. For the clinical-data validation, ten low-dose [$^{18}$F]FDG datasets and one [$^{18}$F]Florbetapir dataset were tested on a model pretrained on data from another tracer. Experiment results show that the proposed PET reconstruction method can generalize robustly across tracer distributions and scanner types, providing an efficient and versatile reconstruction framework for low-dose PET imaging. 2025-07-20T18:25:29Z 11 pages, 12 figures IEEE Trans. Med. Imaging (2026) Fumio Hashimoto Kuang Gong 10.1109/TMI.2026.3659792 http://arxiv.org/abs/2507.21263v2 In-vivo 6D heart motion analysis for emerging self-powered cardiac implants 2025-12-09T09:29:45Z Self-powered intracardiac implant devices show great promise for future clinical applications due to their extended operational lifespan and the potential to reduce the need for high-risk repeat surgeries. This study investigates the feasibility of harvesting energy from cardiac motion through in vivo testing of intracardiac devices. Comprehensive three-dimensional translational and rotational cardiac motions are captured in a porcine model using a miniaturized 9-degree-of-freedom motion sensor implanted at six strategic epicardial sites. Kinematic criteria are developed to evaluate the energy harvesting potential of each implant site based on the available kinetic energy, acceleration, and jerk factors. The recorded heart motion signals are analyzed and applied to a conceptual energy harvester proposed to identify the optimal implant site. The results reveal that the left ventricular apex emerges as a preferable site for energy harvesting, particularly at moderate heart rates. These findings offer valuable insights into optimizing self-powered intracardiac implants, reducing dependency on battery replacements, and enhancing long-term patient safety. 2025-07-28T18:31:19Z Annals of Biomedical Engineering (2026) Milad Hasani John Huber Benedict Kjaergaard Tomas Zaremba Alireza Rezania Sam Riahi 10.1007/s10439-026-04037-5 http://arxiv.org/abs/2512.08014v1 Robust External-Beam Calibration of Plastic Scintillation Detectors for In-Vivo Dosimetry in HDR Brachytherapy 2025-12-08T20:12:51Z Purpose: HDR brachytherapy is a widely adopted modality for cancer treatment. However, it is not free from error and uncertainty. In-vivo dosimetry (IVD) is the only technique that can confirm correct dose delivery. This study details and validates a calibration method for Plastic Scintillation Detector (PSD), bypassing dose gradient and positioning issues in brachytherapy calibration. Methods: The PRB-0057 PSD (Medscint, Canada) was calibrated, 1x1 mm scintillating fiber coupled to a 20 m Eska GH-4001 clear optical fiber (Mitsubishi Rayon, Japan). The fiber is connected to the Hyperscint-RP200 research platform for signal collection. Hyperspectral calibration was performed at a LINAC with a 6 MV beam, enabling removal of stem effects before brachytherapy measurements. For validation, an Iridium-192 Flexisource (Elekta Brachy, The Netherlands), Sk=29447U, was used in a motorized IBA-Blue-Phantom2 water tank (48x48x41cm3). Dose rates were measured at 10 Hz along the source z-axis at a fixed transverse distance of 1.2+/-0.05 cm in 0.2 cm steps. Relative difference (RD) between measured and TG-43U1 dose rates was assessed. A detailed uncertainty budget was associated with brachytherapy measurements. Results: Comparison shows good agreement with RD around 2.5 $\%$ at 1.2 cm, corresponding to positional uncertainties of <0.15 mm. At greater depths up to 8 cm, RDs increase to about 5 $\%$, mainly due to reduced light yield. Uncertainties depend on the source-detector distance, ranging from 3.81 to 6.39 $\%$ (k=1) over the explored range. Conclusions: Results confirm the PSD calibration effectiveness using a 6MV external beam with hyperspectral technique. Uncertainties close to the source align with positional errors and are dominated by reduced PSD sensitivity at larger distances. The study underlined the intrinsic limitation of IVD in the face of known uncertainties. 2025-12-08T20:12:51Z 22 pages, 6 figures, To be submitted in Phys. Med. Biol Chahrazed Ghannoudi Marie-Claude Lavallee Benjamin Cote Luc Beaulieu http://arxiv.org/abs/2312.13976v4 Anatomical basis of sex differences in the electrocardiogram identified by three-dimensional torso-heart imaging reconstruction pipeline 2025-12-08T15:30:11Z The electrocardiogram (ECG) is used for diagnosis and risk stratification in myocardial infarction (MI). Women have a higher incidence of missed MI diagnosis and complications following infarction, and to address this we aim to provide quantitative information on sex-differences in ECG and torso-ventricular anatomical features and their interdependence. A novel computational automated pipeline is presented enabling the three-dimensional reconstruction of torso-ventricular anatomies for 425 post-MI subjects and 1051 healthy controls from UK Biobank clinical images. Regression models were created relating torso-ventricular and ECG parameters. We found that female hearts were positioned more posteriorly and superiorly than male, and in MI hearts were oriented more horizontally, especially for women. Post-MI women exhibited less QRS prolongation, requiring 27% more prolongation than men to exceed 120ms. Only half of the sex difference in QRS duration was associated with smaller female cavities. Lower STj amplitude in women was striking, associated with smaller ventricles, but also more superior and posterior cardiac position. Post-MI, T wave amplitude and R axis deviations were more strongly associated with posterior and horizontal cardiac positioning in women than in men. Our study highlights the need to quantify sex differences in anatomical features, their implications in ECG interpretation, and the application of clinical ECG thresholds in post-MI. 2023-12-21T16:03:56Z Revision following reviewer comments Hannah J. Smith Blanca Rodriguez Yuling Sang Marcel Beetz Robin P. Choudhury Vicente Grau Abhirup Banerjee http://arxiv.org/abs/2512.07598v1 Microfluidic gratings for X-ray Phase Contrast Imaging 2025-12-08T14:46:40Z Fabrication of X-ray gratings has surged in the last two decades thanks to their vast employment in X-ray Phase Contrast Imaging, an imaging technique able to boost X-ray sensitivity to detect otherwise invisible details. These high aspect ratio devices are usually fabricated by complex, costly, multi-step processes that limit their size and volume scaling. These steps commonly involve UV or X-ray lithography, semiconductor selective etching and high-Z metal plating, usually Au, which require expensive tools and materials. Here we present a proof-of-concept fabrication via soft lithography and Hg infusion of microfluidic X-ray absorption gratings and their performance in biomedical imaging. Such fabrication technique requires fewer, less expensive, and more scalable processes using alternative and more sustainable materials, while showing comparable visibility with their conventional Au-based, solid equivalent. Our results constitute a promising shift in X-ray optics fabrication that could significantly lower barriers to commercialization and accelerate the practical deployment of X-ray Phase Contrast Imaging. 2025-12-08T14:46:40Z 13 pages, 4 figures Alessandro Rossi Francesco Coccimiglio Antonio Ferraro Tiziana Ritacco Alberto Astolfo Michele Giocondo Vincenzo Formoso Raffaele Giuseppe Agostino Francesco Iacoviello Ioannis Papakonstantinou Alessandro Olivo