https://arxiv.org/api/Dn8H9bkdTQwGLmjEZoLPSWKijJs 2026-05-16T00:29:18Z 8899 15 15 http://arxiv.org/abs/2605.09267v1 Moving MRI: Imaging a moving body with a moving magnet 2026-05-10T02:25:22Z Current magnetic resonance imaging (MRI) requires the subject to remain stationary to limit motion artifacts and avoid unwanted field-induced brain stimulation. However, imaging during large-scale motion could enable studies in which motion itself is central. One example is the study of brain networks involved in vestibular function, which senses head motion. Here, we demonstrate Moving MRI (mMRI), a system that enables imaging during large-scale motion by moving the subject and scanner together to minimize relative motion. We implemented a proof-of-concept platform using a compact, cryogen-free superconducting magnet mounted on a pneumatically actuated tilt mechanism that moves the magnet, gradients, and RF coil as a unit during scanning. Phantom and in vivo rat brain scans were acquired during repetitive tilting. We characterized artifacts arising from tilt-induced field shifts and residual subject-scanner motion, and partially reduced these effects. mMRI enables imaging during large-scale movement and may broaden access to naturalistic vestibular paradigms while providing a foundation for future human systems. 2026-05-10T02:25:22Z Jingting Yao Artan Kaso Nikhil Patel Yin-Ching Iris Chen Andre van der Kouwe Daniel M. Merfeld Jerome L. Ackerman http://arxiv.org/abs/2605.09174v1 Ortho-Positronium Three-Photon Decays: Physics Constraints and a Closed-Form Energy Method for Annihilation Vertex Reconstruction 2026-05-09T21:25:15Z We examine the physical foundations of orthopositronium three-photon decay in the context of annihilation vertex reconstruction, focusing on how energy-momentum conservation constrains the space of physically admissible solutions. Finally, we provide a closed-form analytical derivation of an energy-based vertex reconstruction algorithm. 2026-05-09T21:25:15Z 13 pages, 4 figures L. Raczyński W. Krzemień A. Coussat M. Bała B. C. Hiesmayr K. Klimaszewski M. Obara R. Y. Shopa http://arxiv.org/abs/2605.09091v1 Combined Diffusion-Relaxation MRI to Assess Muscle Microstructure and Composition 2026-05-09T17:54:58Z Quantifying muscle tissue properties is crucial for understanding pathophysiological changes occurring in skeletal muscle (SM). In particular, T2 relaxation and diffusion MRI (dMRI) are promising techniques. However, typical methods measure T2 and diffusion separately, making them less specific to microstructure than emerging combined diffusion-relaxation techniques. Here we demonstrate a combined diffusion-relaxation MRI approach for disentangling T2 and diffusivity properties in SM. A diffusion-relaxation acquisition was implemented on a 3 T scanner, combining six b-values and four echo times within a 12-min single-slice protocol. Five healthy participants were enrolled. Data were analysed with six microstructural diffusion and diffusion-relaxation models. Mean parameter values were extracted from manually segmented calf muscles. Models neglecting T2 relaxation showed strong TE dependence: mean diffusivity (MD) decreased by up to 47\%, fractional anisotropy (FA) increased by up to 75\%, and vascular fraction fv increased by up to 297\% when TE increased from 50 to 90 ms. Diffusion-relaxation models produced TE-independent estimates. Tissue and vascular relaxation times ranged 31-36 ms T2t and 66-86 ms T2v, respectively. Simulations confirmed improved accuracy for fv estimation (r=0.95; RMSE=0.03) and reduced TE-related bias. Combined diffusion-relaxation MRI provides robust, TE-independent estimates of muscle microstructural and perfusion-related biomarkers. The quantitative improvements observed - particularly in the estimation of fv - show its potential to provide non-invasive biomarkers for the assessment of muscle physiology, exercise adaptation, rehabilitation, and neuromuscular pathology. 2026-05-09T17:54:58Z Matteo Figini Paddy J. Slator Valeria E. Contarino Eleftheria Panagiotaki Giovanna Rizzo Alfonso Mastropietro http://arxiv.org/abs/2601.05717v2 Inclusion of Inter-crystal Scattering in PET: Analytical Models and Dedicated Reconstruction 2026-05-09T17:18:45Z Inter-crystal scattering (ICS) in Positron Emission Tomography (PET) is commonly regarded as a degradation effect that might compromise the image spatial resolution. In parallel, the inclusion of ICS events has also been recognized as a potential approach to increase PET sensitivity, which could be especially beneficial in scenarios where the latter is a limiting factor, such as very small animal imaging. Several methods for the recovery of ICS events have been proposed, many of which aim to locate the first interaction, i.e., the Compton scattering site, usually limited by their success rate, computational burden or data and training dependency. Conversely, this work proposes a physics-based model for ICS events, leading to analytical expressions of the sensitivity image and the system matrix (required by statistical reconstruction algorithms), without the need to identify the original line of response. After validating the model, the work shows how ICS events can be integrated into a joint image reconstruction algorithm (based on list-mode MLEM) together with conventional PET events, for which dedicated analytical models are also developed. To assess the performance of the proposed approach, Monte-Carlo simulated and experimental data of an image quality phantom were obtained with the MERMAID small-fish PET scanner prototype. Both simulation and experimental results indicate that, while slightly decreasing the recovery coefficient values, the inclusion of ICS clearly reduces statistical noise and improves uniformity. 2026-01-09T11:01:05Z This article has been accepted for publication in IEEE Transactions on Radiation and Plasma Medical Sciences. This is the author's version which has not been fully edited and content may change prior to final publication. Citation information: DOI 10.1109/TRPMS.2026.3691192 Jorge Roser Hong Phuc Vo Rebecca Kantorek Steven Seeger Magdalena Rafecas 10.1109/TRPMS.2026.3691192 http://arxiv.org/abs/2605.08957v1 Bilateral breast gradient insert prototype for strong diffusion encoding at 3T 2026-05-09T13:53:15Z Purpose: Diffusion MRI has shown promise for breast cancer screening, lesion characterization,and treatment response monitoring without contrast agents, but further translation is constraint by the gradient performance of conventional systems. The aim of this work is to develop a single-axis high performance bilateral plug-and-play breast gradient insert to enable strong-gradient diffusion MRI. Methods: An in-house breast gradient insert and bed-tabletop was constructed entirely from commercially available materials, providing a cost-effective solution compatible with existing MRI systems. Its wiring pattern was optimized for torque and force balancing, power dissipation, and target field performance. Evaluation included gradient field characterization, peripheral nerve stimulation simulation verification, and temperature and eddy current assessment. The setup was used for imaging of a diffusion phantom based on soy lecithin across a range of b-values. Results: Gradient efficiency reached 2.8 mT/m/A, enabling local strengths up to 1850 mT/m (660 A). No peripheral nerve stimulation was observed during tests on five healthy volunteers. Eddy currents were successfully characterized employed in standard correction methods. Imaging showed the feasibility of $b = 10 000 s/mm^2$ acquisitions at TE = 78 ms versus 161 ms with scanner gradients. Conclusion: This work demonstrates a dedicated bilateral breast gradient insert for safe and feasible strong-gradient breast diffusion MRI, and represents a first step toward dedicated hardware for breast cancer detection and characterization without contrast agents. 2026-05-09T13:53:15Z Gerrit Arends Edwin Versteeg Feng Jia Dennis Klomp Maxim Zaitsev Sebastian Littin Chantal Tax http://arxiv.org/abs/2605.12544v1 Dual-Correction Physics-Informed Neural Networks for Hemodynamic Reconstruction from Sparse Data 2026-05-09T09:32:51Z Quantifying hemodynamics in the curved segments of the intracranial internal carotid artery is a core challenge in diagnosing vascular stenosis. Conventional full-field imaging, such as 4D Flow MRI, is costly and difficult to widely promote. Meanwhile, reconstructing full-field fluid information from easily accessible and non-invasive sparse measurement data (such as transcranial Doppler ultrasound/computed tomography angiography) is essentially a highly challenging ill-posed inverse problem. To overcome the severe optimization difficulties and generalization failures of conventional physics-informed neural networks (PINNs) in highly tortuous geometries, we propose a dual-correction physics-informed neural network (DCP-INN) framework taking into account a causal decoupling strategy. The proposed DCP-INN model utilizes a diamond-shaped main network to capture low-frequency trends in physical evolution, and employs a parallel wide-deep correction network to compensate for high-frequency residuals resulting from complex geometric shapes. Furthermore, the framework introduces a high-order physical loss function based on Taylor expansion to enhance local continuity under extremely sparse data constraints. To validate the proposed method, we performed computational evaluations on realistic vascular geometries with significant tortuosity. The results demonstrate that the method effectively mitigates optimization challenges and significantly reduces flow field reconstruction error. This study not only achieves physically credible and robust flow field reconstruction in complex morphologies but also provides a highly promising algorithmic foundation for building low-cost, high-resolution personalized cardiovascular digital twins in future. 2026-05-09T09:32:51Z 10 pages, 5 figures Jingtai Song Qinsheng Zhu Xiaodong Xing Yufeng Tang Zhiyun Zhang Xianwen Zhang Hao Wu http://arxiv.org/abs/2503.17459v2 Real-time diffuse correlation spectroscopy with a chip-based correlator for measuring human cerebral blood flow and brain function 2026-05-09T07:35:54Z Diffuse correlation spectroscopy (DCS) is a noninvasive optical technique that probes microvascular blood flow in deep tissues. Here, we present and validate a new on-chip hardware correlator for high-speed DCS measurements. The correlator is embedded in a custom-built 512 x 512 single-photon avalanche diode (SPAD) array named ATLAS, which computes intensity autocorrelation functions directly on-chip at a sampling rate of 116 Hz - the fastest DCS acquisition reported to date. Unlike conventional DCS systems that suffer from low light throughput and therefore cannot resolve cardiac pulsations at source-detector separations (rho) beyond 30 mm, our massively parallel on-chip architecture computes autocorrelations within each macropixel, eliminating the data-throughput bottleneck. This enables high-SNR, real-time detection of pulsatile blood flow even at rho = 50 mm on the human forehead. In phantom experiments at rho = 25 mm, ATLAS-DCS achieves a 12-fold improvement in signal-to-noise ratio over a conventional single-channel DCS instrument while operating at 116 Hz. In human subjects, we resolve functional hyperemia during a mental arithmetic task at rho = 30 mm. Furthermore, we integrate ATLAS DCS with a frequency-domain near-infrared spectroscopy (FD-NIRS) module, enabling simultaneous monitoring of blood flow and tissue oxygenation. With this combined system, we can concurrently resolve core hemodynamic parameters. The on-chip parallelized DCS design substantially improves detection speed, depth sensitivity, and real-time capability, paving the way for wearable, high-speed cerebral blood flow monitoring in both clinical and research settings. 2025-03-21T18:13:27Z Quan Wang Yuanyuan Hua Chenxu Li Zhizheng Yuan Jing Wang Ahmet T. Erdogan Hanqing Chen Xunting Huang Maciej Wojtkiewicz Alistair Gorman Mingliang Pan Yuanzhe Zhang Yining Wang Neil Finlayson Renzhe Bi Robert K. Henderson Zhen Yuan David Day-Uei Li http://arxiv.org/abs/2605.08711v1 Automated Optical Density Normalization for Myelin Quantification: Cross-Modal Validation with 7T Ex Vivo MRI 2026-05-09T05:46:41Z White matter hyperintensities (WMH) are bright regions on T2-weighted magnetic resonance imaging (MRI) scans and are associated with cerebrovascular pathology and neurodegeneration, including myelin loss. While Luxol Fast Blue histopathology provides visualization of myelin integrity, quantitative analysis requires measuring Optical Density as a proxy for myelin concentration. However, differences in laboratory protocols and tissue processing introduce staining variability that acts as systematic noise, obscuring the biological signal and preventing consistent comparison across histology runs. To address this, we developed an automated pipeline that identifies reference (non-pathologic) regions in whole-slide images to compute normalized Optical Density heatmaps. We validated this approach through two complementary evaluations: (1) comparison against expert ratings of myelin loss severity, and (2) cross-modal spatial comparison with co-registered 7T ex vivo MRI for voxel-wise evaluation within white matter regions. The pipeline's reference selection showed strong concordance with expert-identified reference regions, and normalized Optical Density demonstrated a substantially stronger correlation with MRI signal intensity than raw measurements. This correlation persisted within WMH, confirming that the pipeline captures continuous myelin pathology rather than merely the presence or absence of myelin loss contrast. By mitigating staining artifacts, this pipeline provides a robust, validated framework for quantitative cross-modal comparison, establishing a critical methodological foundation for future translation to in vivo myelin mapping and biomarker discovery. 2026-05-09T05:46:41Z 10 Pages, accepted at MICCAI 2026 Zahra Khodakarami Sheina Emrani Pulkit Khandelwal Chinmayee Athalye Amanda Denning Winifred Trotman Lisa M Levorse Eric Teunissen-Bermeo Hamsanandini Radhakrishnan Daniel Ohm Christophe Olm Noah Capp Ranjit Ittyerah Karthik Prabhakaran John A. Detre Sandhitsu R. Das David A. Wolk Corey T McMillan Gabor Mizsei M. Dylan Tisdall David J Irwin John L. Robinson Edward B Lee Paul A. Yushkevich http://arxiv.org/abs/2605.08517v1 A Deep Risk Estimator for Known Operator Learning 2026-05-08T21:58:30Z We describe an approach for estimating the statistical risk of deep networks that contain a mix of learned and known operators. Building on the maximal training error bounds previously established for known operator learning, we derive a deep risk estimator that connects the expected error of a layered network to the size of the training sample. The estimator decomposes the total risk into a sum over learned layers; every known operator contributes zero to this sum, while every learned layer adds an approximation term inspired by Barron's classic work and an estimation term that decreases with the number of training samples. We are able to show that the bound shrinks whenever a learned layer is replaced by a known operator and that the corresponding sample requirement scales with the number of trainable parameters of the layer that is replaced. As an application, we use computed tomography as an example and compare an operator-aware filtered backprojection network with a fully connected substitute that collapses the entire reconstruction pipeline into a single learned dense matrix. The predicted parameter ratio coincides with the structural sparsity that the analytic decomposition into a circulant filter and a sparse backprojection exposes. We confirm the predicted scaling on CPU at small image scale and on GPU at medium image scale, all on the same scaling law. Beyond CT reconstruction, the estimator applies to physics-informed neural networks that hardcode a known physical operation in its architecture, and we expect the result to be of interest for a broad community working on operator-aware deep learning. Calibrating the per-layer constants on each sweep yields a bound that tracks the empirical test MSE within a factor of two at every training-set size, so the estimator can be inverted to predict how many training samples are required to reach a target error. 2026-05-08T21:58:30Z In Review Andreas Maier Md Hasan Paulina Conrad Paula Andrea Perez-Toro http://arxiv.org/abs/2605.08403v1 UWB-Fat: Non-Intrusive Body Fat Measurement Using Commodity Ultra-Wideband Radar 2026-05-08T19:10:20Z Body fat percentage and its spatial distribution are clinically important health indicators. However, existing measurement methods often impose a tradeoff between accuracy and accessibility. Clinical-grade techniques, such as Dual-Energy X-ray Absorptiometry (DEXA) and hydrostatic weighing, provide accurate measurements but require specialized equipment and trained operators, making them difficult to access and unsuitable for everyday use. In contrast, consumer-level methods, such as Bioelectrical Impedance Analysis (BIA) smart scales and skinfold calipers, are more accessible but typically provide only coarse-grained estimates, are prone to user error, or require intrusive physical contact. In this work, we present UWB-Fat, the first system that leverages commodity ultra-wideband (UWB) radar to enable non-intrusive, accessible, and accurate caliper-equivalent skinfold thickness estimation, serving as a convenient replacement for the skinfold caliper. UWB-Fat collects UWB signal at specified body sites non-intrusively without operator assistance. It extracts body-composition-related features from UWB signals by exploiting dielectric contrasts among skin, fat, and muscle tissues. Then, it uses a physics-inspired model to estimate site-specific skinfold thickness. We evaluate UWB-Fat on 15 participants, achieving a root mean square error of 0.63~mm for pooled-site subcutaneous fat thickness. These results highlight the potential of UWB-Fat to support low-cost, self-administered, and everyday body fat monitoring. 2026-05-08T19:10:20Z Haotang Li Yili Ren Zhenyu Qi Sen He Kebin Peng Sheng Tan Bo Liu Jiyue Zhao Zi Wang http://arxiv.org/abs/2605.06558v2 John Equation Constraints for the 3D X-ray Transform under a Cylindrical-Spherical Mixed Parameterization: Theoretical Derivation, Experimental Validation, and Application Analysis 2026-05-08T17:03:05Z The John equation serves as the mathematical foundation of the X-ray transform, describing the intrinsic compatibility conditions that projection data must satisfy. In this paper, within three-dimensional (3D) Euclidean space, an innovative mixed parameterization scheme is adopted: the source point is represented using cylindrical coordinates a=(s cosθ,s sinθ,z_0), and the ray direction is represented using spherical coordinates d=\{rho}(-cos\{beta}sinα,cos\{beta}cosα,sin\{beta}). The specific form of the John equation under this geometric parameterization is systematically derived. Through detailed partial differential operator transformations, application of -1 homogeneity, and algebraic simplification, a complete system of constraint equations is obtained. In particular, under the special configurations where the ray direction is perpendicular to the radial direction of the source point in the horizontal plane (i.e., the so-called alignment condition:α = θ) and the ray has no tilt (\{beta} = 0), the constraint equations simplify to differential relations with clear physical meanings. This paper not only establishes a bridge between abstract mathematical theory and concrete imaging geometry, but also provides rigorous mathematical tools for data consistency verification, geometric parameter calibration, and incomplete-data reconstruction in 3D Computed Tomography (CT) systems. The research results are of great significance for advancing the mathematical theory and practical applications of CT imaging. 2026-05-07T16:51:51Z Shaojie Tang Zhiwei Qiao Xuanqin Mou http://arxiv.org/abs/2605.07966v1 Bubble jetting in acoustic microdroplet vaporization 2026-05-08T16:29:29Z Acoustic droplet vaporization denotes the phase-change of micron- and sub-micron-sized droplets upon the application of high-amplitude ultrasound. The asymmetric collapse of the incepted vapor bubbles within the droplets can give rise to high-speed liquid microjets. Here, we describe acoustically-driven and bubble-pair jetting arising within the vaporizing droplet, observed experimentally with ultra-high-speed imaging at the microscale. The existence of complex pressure fields due to the continued acoustic wave-droplet interaction and the nucleation of multiple bubbles within the droplet leads to rich dynamics, with the jets presenting behavioral self-similarity to millimetric bubbles under comparable conditions. Evaporative instabilities that develop during bubble growth impede jet formation during bubble collapse. Furthermore, the ability of the jets to pierce the droplet interface and penetrate into the surrounding fluid is discussed. These powerful microjets could be harnessed to induce cell permeabilization for targeted drug delivery and treatment of cancerous tissue. 2026-05-08T16:29:29Z Anunay Prasanna Samuele Fiorini Gazendra Shakya Outi Supponen http://arxiv.org/abs/2605.08264v1 Indirect Detection of Lactate Through Voltammetry Using Glassy Carbon Microelectrodes 2026-05-08T00:07:53Z Glassy carbon (GC) microelectrodes are increasingly being used for voltametric detection of electroactive neurotransmitters such as dopamine and serotonin. However, non-electroactive molecules including lactate, glutamate, and gamma-aminobutyric acid (GABA) cannot be directly detected using conventional voltammetry without surface functionalization. In this study, lactate oxidase was immobilized within a chitosan matrix on lithographically patterned GC microelectrodes to enable indirect detection of lactate via enzymatic generation of hydrogen peroxide, an electroactive byproduct. The resulting hydrogen peroxide was detected using fast-scan cyclic voltammetry (FSCV), enabling indirect in vitro detection of lactate at concentrations as low as 10 nM. The functionalized GC microelectrodes were integrated into a four channel array on a 1.6 cm flexible neural probe with potential for in vivo applications. Surface morphology and bonding interactions were characterized using scanning electron microscopy (SEM) and Fourier transform infrared (FTIR) spectroscopy. FTIR analysis confirmed successful chitosan deposition through characteristic O-H, N-H, amide, and C-O stretching bands. Hydrogen peroxide detection was concentration-dependent, while lactate detection exhibited early saturation consistent with enzyme-limited kinetics. These results demonstrate a mechanically robust GC microelectrode platform for nanomolar-level indirect lactate sensing and provide insight into the reaction-diffusion coupling governing enzyme-based electrochemical detection. 2026-05-08T00:07:53Z 14 pages; 10 figures Amish Rohtagia Elisa Barts Neharika Ravichandran Sandra Lara Galindoa Surabhi Nimbalkara Mantra Mittal Samantha Omer Sam Kassegnea http://arxiv.org/abs/2605.06820v1 Overcoming data scarcity through multi-center federated learning for organs-at-risk segmentation in pediatric upper abdominal radiotherapy 2026-05-07T18:21:43Z Deep learning-based organs/structures-at-risk(OARs) auto-contouring models can improve radiotherapy workflows, but models trained on adult data often underperform in pediatric patients. Developing robust pediatric-specific models is hindered by data scarcity and fragmentation across centers. Federated learning (FL) enables privacy-preserving collaborative training without the need for data sharing. We evaluated the feasibility and performance of FL for developing pediatric-specific OAR segmentation models across two European medical centers. Computed tomography (CT) images from pediatric patients from Utrecht and Heidelberg with a renal tumor or abdominal neuroblastoma were retrospectively collected and locally processed. An nnU-Net-based framework segmented 19 OARs using local and FL schemes. FL was implemented with secure weight exchange on a cloud storage across institutional firewalls. Performance was assessed using the Dice similarity coefficient (DSC), 95th percentile Hausdorff distance, and mean surface distance. Robustness to patient orientation, false-positive segmentation of surgically removed kidneys, and failure cases were identified. A total of 310 postoperative CTs from 272 patients (105 renal tumors, 167 neuroblastomas) were included. Local models performed well on their respective center data but showed significantly reduced cross-center performance for four to seven of the nine evaluated OARs (DSC). In contrast, the FL model matched local performance for at least seven of nine OARs and achieved the best cross-center results across three metrics, with DSC gains of 0.003-0.007 over local models. FL also maintained stable performance across patient orientations and reduced false-positive kidney segmentations. Real-world FL improves cross-center robustness of CT-based OAR segmentation models in pediatric upper abdominal tumors. 2026-05-07T18:21:43Z Mianyong Ding Maximilian Knoll Semi Harrabi Martine van Grotel Annemieke S. Littooij Max van Noesel Jens-Peter Schenk Marry M. van den Heuvel-Eibrink Geert O. Janssens Matteo Maspero http://arxiv.org/abs/2602.12805v2 A Wavefield Correlation Approach to Improve Sound Speed Estimation in Ultrasound Autofocusing 2026-05-07T16:54:40Z In pulse-echo ultrasound, aberration often degrades image quality when beamforming does not account for wavefront distortions. To address this issue, local sound speed estimators have been developed in the past decade for distributed aberration correction. Recently, methods based on iterative optimization have improved sound speed accuracy with respect to earlier approaches. However, the accuracy of these newer methods is limited by media with reverberation clutter and by the straight-ray model of wave propagation. To address these challenges, we propose using wavefield correlation (WFC) beamforming when performing sound speed optimization. WFC, an ultrasound adaptation of reverse time migration, correlates simulated forward-propagated transmit wavefields and backwards-propagated receive wavefields in order to reconstruct images. This process more accurately models wave propagation in heterogeneous media and can decrease diffuse clutter due to its spatiotemporal matched filtering effect. We implement herein a WFC beamformer using an auto-differentiation software and estimate the sound speed map by optimizing a regularized common-midpoint phase focusing criterion using gradient descent. This approach is compared to a previous method relying on delay and sum (DAS) with straight-ray time delay calculations on a variety of simulated, phantom, and in vivo data with large sound speed variations and clutter. Results show that using WFC decreases sound speed estimation error, leading to improvements in resolution and contrast in the corrected image. In particular, these promising results have potential to improve pulse-echo imaging for challenging clinical scenarios. 2026-02-13T10:40:20Z Louise Zhuang Samuel Beuret Ben Frey Saachi Munot Walter Simson Dongwoon Hyun Jeremy J. Dahl