https://arxiv.org/api/J04L5DxuR/3Sd6TotvK44rLICEU 2026-06-27T11:02:53Z 9008 1410 15 http://arxiv.org/abs/2504.14637v1 Transferred plasma catheter for endotherapeutic applications: a parametric study of guided streamers dynamics 2025-04-20T14:29:57Z Non-thermal atmospheric pressure plasma jets (APPJs) are increasingly used in biomedical applications due to their low temperatures and ability to generate reactive oxygen and nitrogen species (RONS), making them suitable for sensitive environments like medical therapies. The transferred plasma catheter (TPC), a variant of APPJ, shows promise for endoscopic applications but requires precise control of plasma dynamics in confined spaces to ensure safety and efficacy. Despite extensive studies on guided streamers in traditional APPJs, there is limited understanding of streamer behavior in TPC configurations, particularly in challenging scenarios involving grounded metallic surfaces. This study examines the spatiotemporal dynamics of guided streamers generated by TPCs under varying gap distances to establish a robust framework for safe and effective plasma delivery in endoscopic settings. Combining electrical and optical diagnostics, the study characterizes streamer propagation, electric field profiles, and plasma-induced currents in a helium-driven TPC delivering cold plasma to a grounded metal target across gaps of 2 to 18 mm. Results show that streamers maintain charge stability and effectively interact with the target for gap distances below 12 mm, producing significant therapeutic currents. Beyond this threshold, propagation deteriorates due to recombination and reduced electric field intensity. For shorter gaps, counter-propagating waves and secondary streamer interactions are observed, while larger gaps lead to charge dissipation and reduced efficacy. These findings highlight the importance of optimizing gap distances for plasma-assisted endoscopic procedures and demonstrate the TPC's robustness in adverse conditions. 2025-04-20T14:29:57Z Physics of Plasma, Vol. 32, Issue 4 (2025) M. Soulier T. Vacek K. Geraud T. Dufour 10.1063/5.0254402 http://arxiv.org/abs/2502.13147v2 Impact of Optic Nerve Tortuosity, Globe Proptosis, and Size on Retinal Ganglion Cell Thickness Across General, Glaucoma, and Myopic Populations: Insights from the UK Biobank 2025-04-19T20:19:02Z Purpose: To investigate the impact of optic nerve tortuosity (ONT), and the interaction of globe proptosis and globe size on retinal ganglion cell (RGC) thickness, using Retinal Nerve Fiber Layer (RNFL) thickness, across general, glaucoma, and myopic populations. Methods: We analyzed 17,940 eyes from the UKBiobank cohort (ID 76442), including 72 glaucoma and 2475 myopic eyes. AI models segmented structures from 3D optical coherence tomography (OCT) scans and magnetic resonance images (MRI). RNFL thickness was derived from OCT scans and corrected for ocular magnification, was derived from OCT. From MRIs, we extracted: ONT, globe proptosis, axial length, and a novel interzygomatic line-to-posterior pole (ILPP) distance, a composite marker of globe proptosis and size. GEE models assessed associations between orbital and retinal features across all populations. Results: Segmentation models achieved Dice coefficients over 0.94 for both MRI and OCT. RNFL thickness was positively correlated with both ONT and ILPP distance (r = 0.065, p < 0.001, and r = 0.206, p < 0.001 respectively). The same was true for glaucoma (r = 0.040, p = 0.74, and r = 0.224, p = 0.059), and for myopia (r = 0.069, p < 0.001, and r = 0.100, p < 0.0001). GEE models revealed straighter optic nerves and shorter ILPP distance as predictive of thinner RNFL in all populations. Conclusions: This study emphasizes the impact of ONT, globe size, and proptosis on retinal health, suggesting RNFL thinning may arise from biomechanical stress due to straighter optic nerves or reduced ILPP distance, particularly in glaucoma or myopia. The novel ILPP metric, integrating globe size and position, shows potential as a biomarker for axonal health. These findings highlight the role of orbit structures in RGC axonal health and warrant further exploration of the biomechanical relationship between the orbit and optic nerve. 2025-01-22T21:24:42Z 40 pages, 6 figures, 4 tables, 1 supplementary material Charis Y. N. Chiang Xiaofei Wang Stuart K. Gardiner Martin Buist Michael J. A. Girard http://arxiv.org/abs/2504.14227v1 Reducing the set of considered scenarios in robust optimization of intensity-modulated proton therapy 2025-04-19T08:25:36Z Robust optimization is a commonly employed method to mitigate uncertainties in the planning of intensity-modulated proton therapy (IMPT). In certain contexts, the large number of uncertainty scenarios makes the robust problem impractically expensive to solve. Recent developments in research on IMPT treatment planning indicate that the number of ideally considered error scenarios may continue to increase. In this paper, we therefore investigate methods that reduce the size of the scenario set considered during the robust optimization. Six cases of patients with non-small cell lung cancer are considered. First, we investigate the existence of an optimal subset of scenarios that needs to be considered during robust optimization, and perform experiments to see if the set can be found in a reasonable time and substitute for the full set of scenarios during robust IMPT optimization. We then consider heuristic methods to estimate this subset or find subsets with similar properties. Specifically, we select a subset of maximal diversity in terms of scenario-specific features such as the dose distributions and function gradients at the initial point. Finally, we consider adversarial methods as an alternative to solving the full robust problem and investigate the impact on computation times. The results indicate that the optimal subset can be well approximated by solving the robust IMPT problem with conventional methods. Of the methods designed to approximate it within a practically useful time frame, the results of the diversity-maximization methods indicate that they may perform better than a manual selection of scenarios based on the patient geometry. In addition, the adversarial approaches decreased the computation time by at least half compared to the conventional approach. 2025-04-19T08:25:36Z Ivar Bengtsson http://arxiv.org/abs/2410.03802v3 Mesh-Informed Reduced Order Models for Aneurysm Rupture Risk Prediction 2025-04-17T12:45:16Z The complexity of the cardiovascular system needs to be accurately reproduced in order to promptly acknowledge health conditions; to this aim, advanced multifidelity and multiphysics numerical models are crucial. On one side, Full Order Models (FOMs) deliver accurate hemodynamic assessments, but their high computational demands hinder their real-time clinical application. In contrast, Reduced Order Models (ROMs) provide more efficient yet accurate solutions, essential for personalized healthcare and timely clinical decision-making. In this work, we explore the application of computational fluid dynamics (CFD) in cardiovascular medicine by integrating FOMs with ROMs for predicting the risk of aortic aneurysm growth and rupture. Wall Shear Stress (WSS) and the Oscillatory Shear Index (OSI), sampled at different growth stages of the thoracic aortic aneurysm, are predicted by means of Graph Neural Networks (GNNs). GNNs exploit the natural graph structure of the mesh obtained by the Finite Volume (FV) discretization, taking into account the spatial local information, regardless of the dimension of the input graph. Our experimental validation framework yields promising results, confirming our method as a valid alternative that overcomes the curse of dimensionality. 2024-10-04T09:39:15Z Giuseppe Alessio D'Inverno Saeid Moradizadeh Sajad Salavatidezfouli Pasquale Claudio Africa Gianluigi Rozza http://arxiv.org/abs/2504.12772v1 Artifacts in Photoacoustic Imaging: Origins and Mitigations 2025-04-17T09:13:27Z Photoacoustic imaging (PAI) is rapidly moving from the laboratory to the clinic, increasing the need to understand confounders which might adversely affect patient care. Over the past five years, landmark studies have shown the clinical utility of PAI, leading to regulatory approval of several devices. In this article, we describe the various causes of artifacts in PAI, providing schematic overviews and practical examples, simulated as well as experimental. This work serves two purposes: (1) educating clinical users to identify artifacts, understand their causes, and assess whether their impact, and (2) providing a reference of the limitations of current systems for those working to improve them. We explain how two aspects of PAI systems lead to artifacts: their inability to measure complete data sets, and embedded assumptions during reconstruction. We describe the physics underlying PAI, and propose a classification of the artifacts. The paper concludes by discussing possible advanced mitigation strategies. 2025-04-17T09:13:27Z Max T. Rietberg Janek Gröhl Thomas R. Else Sarah E. Bohndiek Srirang Manohar Ben T. Cox http://arxiv.org/abs/2504.11273v2 Hybrid Compton-PET Imaging for ion-range verification:A Preclinical Study for Proton-, Helium-, and Carbon-Therapy at HIT 2025-04-17T08:19:14Z Enhanced-accuracy ion-range verification in real time shall enable a significant step forward in the use of therapeutic ion beams. Positron-emission tomography (PET) and prompt-gamma imaging (PGI) are two of the most promising and researched methodologies, both of them with their own advantages and challenges. Thus far, both of them have been explored for ion-range verification in an independent way. However, the simultaneous combination of PET and PGI within the same imaging framework may open-up the possibility to exploit more efficiently all radiative emissions excited in the tissue by the ion beam. Here we report on the first pre-clinical implementation of an hybrid PET-PGI imaging system, hereby exploring its performance over several ion-beam species (H, He and C), energies (55 MeV to 275 MeV) and intensities (10$^7$-10$^9$ ions/spot), which are representative of clinical conditions. The measurements were carried out using the pencil-beam scanning technique at the synchrotron accelerator of the Heavy Ion Therapy centre in Heidelberg utilizing an array of four Compton cameras in a twofold front-to-front configuration. The results demonstrate that the hybrid PET-PGI technique can be well suited for relatively low energies (55-155 MeV) and beams of protons. On the other hand, for heavier beams of helium and carbon ions at higher energies (155-275 MeV), range monitoring becomes more challenging owing to large backgrounds from additional nuclear processes. The experimental results are well understood on the basis of realistic Monte Carlo (MC) calculations, which show a satisfactory agreement with the measured data. This work can guide further upgrades of the hybrid PET-PGI system towards a clinical implementation of this innovative technique. 2025-04-15T15:14:28Z Javier Balibrea-Correa Jorge Lerendegui-Marco Ion Ladarescu Sergio Morell Carlos Guerrero Teresa Rodríguez-González Maria del Carmen Jiménez-Ramos Jose Manuel Quesada Julia Bauer Stephan Brons César Domingo-Pardo http://arxiv.org/abs/2411.06447v2 Multi-Parameter Molecular MRI Quantification using Physics-Informed Self-Supervised Learning 2025-04-17T05:54:02Z Biophysical model fitting plays a key role in obtaining quantitative parameters from physiological signals and images. However, the model complexity for molecular magnetic resonance imaging (MRI) often translates into excessive computation time, which makes clinical use impractical. Here, we present a generic computational approach for solving the parameter extraction inverse problem posed by ordinary differential equation (ODE) modeling coupled with experimental measurement of the system dynamics. This is achieved by formulating a numerical ODE solver to function as a step-wise analytical one, thereby making it compatible with automatic differentiation-based optimization. This enables efficient gradient-based model fitting, and provides a new approach to parameter quantification based on self-supervised learning from a single data observation. The neural-network-based train-by-fit pipeline was used to quantify semisolid magnetization transfer (MT) and chemical exchange saturation transfer (CEST) amide proton exchange parameters in the human brain, in an in-vivo molecular MRI study (n = 4). The entire pipeline of the first whole brain quantification was completed in 18.3 $\pm$ 8.3 minutes. Reusing the single-subject-trained network for inference in new subjects took 1.0 $\pm$ 0.2 s, to provide results in agreement with literature values and scan-specific fit results. 2024-11-10T12:40:33Z This project was funded by the European Union (ERC, BabyMagnet, project no. 101115639), the Ministry of Innovation, Science and Technology, Israel, and a grant from the Tel Aviv University Center for AI and Data Science (TAD, The Blavatnik AI and Data Science Fund). None of above can be held responsible for views and opinions expressed, which are those of the authors alone Commun Phys 8, 164 (2025) Alex Finkelstein Nikita Vladimirov Moritz Zaiss Or Perlman 10.1038/s42005-025-02063-8 http://arxiv.org/abs/2504.12010v2 Dual-Energy Cone-Beam CT Using Two Orthogonal Projection Views: A Phantom Study 2025-04-17T01:24:47Z This study proposes a novel imaging and reconstruction framework for dual-energy cone-beam CT (DECBCT) using only two orthogonal X-ray projections at different energy levels (2V-DECBCT). The goal is to enable fast and low-dose DE volumetric imaging with high spectral fidelity and structural accuracy, suitable for DECBCT-guided radiation therapy. We introduce a framework for 2V-DECBCT based on physics-informed dual-domain diffusion models. A cycle-domain training strategy is employed to enforce consistency between projection and volume reconstructions through a differentiable physics-informed module. Furthermore, a spectral-consistency loss is introduced to preserve inter-energy contrast during the generative process. The model is trained and evaluated using 4D XCAT phantom data under realistic anatomical motion. The method produces high-fidelity DECBCT volumes from only two views, accurately preserving anatomical boundaries and suppressing artifacts. Subtraction maps computed from the reconstructed energy volumes show strong visual and numerical agreement with ground truth. This work presents the first diffusion model-based framework for 2V-DECBCT reconstruction, demonstrating accurate structural and spectral recovery from extremely sparse inputs. 2025-04-16T12:05:33Z Junbo Peng Tonghe Wang Shaoyan Pan Xiaofeng Yang http://arxiv.org/abs/2504.12531v1 A theoretical framework for flow-compatible reconstruction of heart motion 2025-04-16T23:40:20Z Accurate three-dimensional (3D) reconstruction of cardiac chamber motion from time-resolved medical imaging modalities is of growing interest in both the clinical and biomechanical fields. Despite recent advancement, the cardiac motion reconstruction process remains complex and prone to uncertainties. Moreover, traditional assessments often focus on static comparisons, lacking assurances of dynamic consistency and physical relevance. This work introduces a novel paradigm of flow-compatible motion reconstruction, integrating anatomical imaging with flow data to ensure adherence to fundamental physical principles, such as mass and momentum conservation. The approach is demonstrated in the context of right ventricular motion, utilizing diffeomorphic mappings and multi-slice MRI to achieve dynamically consistent and physically robust reconstructions. Results show that enforcing flow compatibility within the reconstruction process is feasible and enhances the physical realism of the resulting kinematics. 2025-04-16T23:40:20Z Francesco Capuano Yue-Hin Loke Ibrahim Yildiran Laura Olivieri Elias Balaras http://arxiv.org/abs/2504.12438v1 Hybrid artificial intelligence echogenic components-based diagnosis of adnexal masses on ultrasound 2025-04-16T19:12:00Z Background: Adnexal masses are heterogeneous and have varied sonographic presentations, making them difficult to diagnose correctly. Purpose: Our study aimed to develop an innovative hybrid artificial intelligence/computer aided diagnosis (AI/CADx)-based pipeline to distinguish between benign and malignant adnexal masses on ultrasound imaging based upon automatic segmentation and echogenic-based classification. Methods: The retrospective study was conducted on a consecutive dataset of patients with an adnexal mass. There was one image per mass. Mass borders were segmented from the background via a supervised U-net algorithm. Masses were spatially subdivided automatically into their hypo- and hyper-echogenic components by a physics-driven unsupervised clustering algorithm. The dataset was separated by patient into a training/validation set (95 masses; 70%) and an independent held-out test set (41 masses; 30%). Eight component-based radiomic features plus a binary measure of the presence or absence of solid components were used to train a linear discriminant analysis classifier to distinguish between malignant and benign masses. Classification performance was evaluated using the area under the receiver operating characteristic curve (AUC), along with sensitivity, specificity, negative predictive value, positive predictive value, and accuracy at target 95% sensitivity. Results: The cohort included 133 patients with 136 adnexal masses. In distinguishing between malignant and benign masses, the pipeline achieved an AUC of 0.90 [0.84, 0.95] on the training/validation set and 0.93 [0.83, 0.98] on the independent test set. Strong diagnostic performance was observed at the target 95% sensitivity. Conclusions: A novel hybrid AI/CADx echogenic components-based ultrasound imaging pipeline can distinguish between malignant and benign adnexal masses with strong diagnostic performance. 2025-04-16T19:12:00Z 25 pages, 7 figures, 4 tables Roni Yoeli-Bik Heather M. Whitney Hui Li Agnes Bilecz Jacques S. Abramowicz Li Lan Ryan E. Longman Maryellen L. Giger Ernst Lengyel 10.1002/mp.17983 http://arxiv.org/abs/2504.11847v1 Ultra-Efficient Kidney Stone Fragment Removal via Spinner-Induced Synergistic Circulation and Spiral Flow 2025-04-16T08:10:16Z Kidney stones can cause severe pain and complications such as chronic kidney disease or kidney failure. Retrograde intrarenal surgery (RIRS), which uses laser lithotripsy to fragment stones for removal via a ureteroscope, is widely adopted due to its safety and effectiveness. However, conventional fragment removal methods using basketing and vacuum-assisted aspiration are inefficient, as they can capture only 1 to 3 fragments (1--3\,mm in size) per pass, often requiring dozens to hundreds of ureteroscope passes during a single procedure to completely remove the fragments. These limitations lead to prolonged procedures and residual fragments that contribute to high recurrence rates. To address these limitations, we present a novel spinner device that enables ultra-efficient fragment removal through spinning-induced localized suction. The spinner generates a three-dimensional spiral and circulating flow field that dislodges and draws fragments into its cavity even from distances over 20\,mm, eliminating the need to chase fragments. It can capture over 60 fragments (0.5--2\,mm) or over 15 larger fragments (2--3\,mm) in a single pass, significantly improving removal efficiency. In this work, the spinner design is optimized via computational fluid dynamics to maximize suction performance. \textit{In vitro} testing demonstrates near 100\% capture rates for up to 60 fragments in a single operation and superior large-distance capture efficacy compared to vacuum-assisted methods. \textit{Ex vivo} testing of the integrated spinner-ureteroscope system in a porcine kidney confirmed its high performance by capturing 45 fragments in just 4 seconds during a single pass and achieving complete fragment clearance within a few passes. 2025-04-16T08:10:16Z Yilong Chang Jasmine Vallejo Yangqing Sun Ruike Renee Zhao http://arxiv.org/abs/2504.11578v1 A Comparative Evaluation of CT Global Noise Calculation Methods for Clinical Image Quality Assessment 2025-04-15T19:55:36Z The CMS quality measure for CT mandates compliance with dose and image quality across 18 exam categories, yet the method for calculating global noise (GN) remains undefined, referencing only the Duke and Wisconsin methods. This ambiguity raises concerns about standardization and clinical applicability. We retrospectively analyzed 719 CT exams across seven protocols, comparing five GN metrics derived from the Duke and Wisconsin approaches. Results showed significant variability across metrics and protocols, with Wisconsin_tissue_mean consistently yielding the highest values. Correlation strength varied by protocol, and mode-based metrics demonstrated higher agreement, suggesting dose dependence. In the Chest PE protocol, GN values differed significantly based on reconstruction kernel. These findings underscore the need for a standardized, clinically meaningful GN metric to ensure consistent image quality evaluation and effective implementation of the CMS measure. 2025-04-15T19:55:36Z Charles M Weaver Gary Ge Alexander Alsalihi Jie Zhang http://arxiv.org/abs/2504.11519v1 FACT: Foundation Model for Assessing Cancer Tissue Margins with Mass Spectrometry 2025-04-15T16:36:03Z Purpose: Accurately classifying tissue margins during cancer surgeries is crucial for ensuring complete tumor removal. Rapid Evaporative Ionization Mass Spectrometry (REIMS), a tool for real-time intraoperative margin assessment, generates spectra that require machine learning models to support clinical decision-making. However, the scarcity of labeled data in surgical contexts presents a significant challenge. This study is the first to develop a foundation model tailored specifically for REIMS data, addressing this limitation and advancing real-time surgical margin assessment. Methods: We propose FACT, a Foundation model for Assessing Cancer Tissue margins. FACT is an adaptation of a foundation model originally designed for text-audio association, pretrained using our proposed supervised contrastive approach based on triplet loss. An ablation study is performed to compare our proposed model against other models and pretraining methods. Results: Our proposed model significantly improves the classification performance, achieving state-of-the-art performance with an AUROC of $82.4\% \pm 0.8$. The results demonstrate the advantage of our proposed pretraining method and selected backbone over the self-supervised and semi-supervised baselines and alternative models. Conclusion: Our findings demonstrate that foundation models, adapted and pretrained using our novel approach, can effectively classify REIMS data even with limited labeled examples. This highlights the viability of foundation models for enhancing real-time surgical margin assessment, particularly in data-scarce clinical environments. 2025-04-15T16:36:03Z International Journal of Computer Assisted Radiology and Surgery (2025) Mohammad Farahmand Amoon Jamzad Fahimeh Fooladgar Laura Connolly Martin Kaufmann Kevin Yi Mi Ren John Rudan Doug McKay Gabor Fichtinger Parvin Mousavi 10.1007/s11548-025-03355-8 http://arxiv.org/abs/2504.11188v1 Clinically Interpretable Survival Risk Stratification in Head and Neck Cancer Using Bayesian Networks and Markov Blankets 2025-04-15T13:47:35Z Purpose: To identify a clinically interpretable subset of survival-relevant features in HN cancer using Bayesian Network (BN) and evaluate its prognostic and causal utility. Methods and Materials: We used the RADCURE dataset, consisting of 3,346 patients with H&N cancer treated with definitive (chemo)radiotherapy. A probabilistic BN was constructed to model dependencies among clinical, anatomical, and treatment variables. The Markov Blanket (MB) of two-year survival (SVy2) was extracted and used to train a logistic regression model. After excluding incomplete cases, a temporal split yielded a train/test (2,174/820) dataset using 2007 as the cutoff year. Model performance was assessed using area under the ROC curve (AUC), C-index, and Kaplan-Meier (KM) survival stratification. Model fit was further evaluated using a log-likelihood ratio (LLR) test. Causal inference was performed using do-calculus interventions on MB variables. Results: The MB of SVy2 included 6 clinically relevant features: ECOG performance status, T-stage, HPV status, disease site, the primary gross tumor volume (GTVp), and treatment modality. The model achieved an AUC of 0.65 and C-index of 0.78 on the test dataset, significantly stratifying patients into high- and low-risk groups (log-rank p < 0.01). Model fit was further supported by a log-likelihood ratio of 70.32 (p < 0.01). Subgroup analyses revealed strong performance in HPV-negative (AUC = 0.69, C-index = 0.76), T4 (AUC = 0.69, C-index = 0.80), and large-GTV (AUC = 0.67, C-index = 0.75) cohorts, each showing significant KM separation. Causal analysis further supported the positive survival impact of ECOG 0, HPV-positive status, and chemoradiation. Conclusions: A compact, MB-derived BN model can robustly stratify survival risk in HN cancer. The model enables explainable prognostication and supports individualized decision-making across key clinical subgroups. 2025-04-15T13:47:35Z 24 pages, 7 figures, 2 tables Keyur D. Shah Ibrahim Chamseddine Xiaohan Yuan Sibo Tian Richard Qiu Jun Zhou Anees Dhabaan Hania Al-Hallaq David S. Yu Harald Paganetti Xiaofeng Yang 10.1016/j.ijrobp.2025.09.063 http://arxiv.org/abs/2412.11593v3 Exploring Offline Pileup Correction to Improve the Accuracy of Microdosimetric Characterization in Clinical Ion Beams 2025-04-15T10:30:35Z Microdosimetry investigates the energy deposition of ionizing radiation at microscopic scales, beyond the assessment capabilities of macroscopic dosimetry. This contributes to an understanding of the biological response in radiobiology, radiation protection and radiotherapy. Microdosimetric pulse height spectra are usually measured using an ionization detector in a pulsed readout mode. This incorporates a charge-sensitive amplifier followed by a shaping network. At high particle rates, the pileup of multiple pulses leads to distortions in the recorded spectra. Especially for gas-based detectors, this is a significant issue, that can be reduced by using solid-state detectors with smaller cross-sectional areas and faster readout speeds. At particle rates typical for ion therapy, however, such devices will also experience pileup. Mitigation techniques often focus on avoiding pileup altogether, while post-processing approaches are rarely investigated. This work explores pileup effects in microdosimetric measurements and presents a stochastic resampling algorithm, allowing for offline simulation and correction of spectra. Initially it was developed for measuring neutron spectra with tissue equivalent proportional counters and is adapted for the use with solid-state microdosimeters in a clinical radiotherapy setting. The algorithm was tested on data acquired with solid-state microdosimeters at the MedAustron ion therapy facility. The successful simulation and reduction of pileup counts is achieved by establishing of a limited number of parameters for a given setup. The presented results illustrate the potential of offline correction methods in situations where a direct pileup-free measurement is currently not practicable. 2024-12-16T09:30:19Z Revision prepared for submission to Physics in Medicine & Biology Matthias Knopf Sandra Barna Daniel Radmanovac Thomas Bergauer Albert Hirtl Giulio Magrin 10.1088/1361-6560/ade6bc