https://arxiv.org/api/jme+Gr1gz87/rKyKiuJm5nB3le8 2026-06-26T15:34:25Z 9008 1125 15 http://arxiv.org/abs/2507.21481v1 Patient-Specific Modeling of Dose-Escalated Proton Beam Therapy for Locally Advanced Pancreatic Cancer 2025-07-29T03:49:43Z Purpose: This study explores the feasibility of dose-escalated proton beam therapy (dPBT) for Locally Advanced Pancreatic Cancer (LAPC) patients by modeling common patient scenarios using current clinically-adopted practices. Methods: Five patient datasets were used as simulation phantoms, each with six tumour sizes, to systematically simulate treatment scenarios typical in LAPC patients. Using the Raystation treatment planning system, robustly-optimised dPBT and stereotactic ablative radiotherapy (SABR) treatment plans were created with a 5 mm margin allowing for intra- and inter-fraction anatomical changes. following clinically-adopted protocols. Safe dose-escalation feasibility is assessed with dose metrics, tumour control (TCP) and normal tissue complication probabilities (NTCP) for average and worst-case intra-fraction motion scenarios. Significance testing was performed using a paired student's t-test. Results: Dose-escalation feasibility is largely dependent on tumour size and proximity to critical structures. Minimal therapeutic benefit was observed for patients with greater than 4.5 cm tumours, however for tumours less than or equal to 4.5 cm dPBT TCPs of 45-90% compared to SABR TCPs of 10-40% (p<0.05). The worst-case scenario dPBT TCP was comparable to SABR. Hypofractioned dPBT further improved this result to greater than 90% (p<0.05) for tumours less than or equal to 4.5 cm. Conclusion: Safe dPBT is feasible for patients with targets up to the median size and see a significant therapeutic benefit compared to the current standard of care in SABR. A patient-specific approach should be taken based on tumour size and surrounding anatomy. 2025-07-29T03:49:43Z M. A. McIntyre J. Midson P. Wilson P Gorayski C. E. Hsieh S. W. Wu E. Bezak http://arxiv.org/abs/2507.21427v1 Depth-resolved polarisation sensitive optical coherence tomography reveals the complex microanatomical response of cartilage to compression 2025-07-29T01:35:09Z Conventional methods for analysing cartilage microstructure under mechanical loading are largely destructive. In this work, we evaluate the efficacy of using depth-resolved polarisation sensitive optical coherence tomography (PS-OCT) to study the cartilage morphological response to compression. We show that depth-resolved PS-OCT reveals the microstructure of cartilage under load, and it can do so non-destructively, opening significant possibilities for enhanced clinical assessment of cartilage health by detecting deviance from normal load-bearing behaviour. 2025-07-29T01:35:09Z 23 pages, 11 figures Darven Murali Tharan The University of Auckland, Department of Physics, New Zealand The Dodd Walls Centre for Quantum and Photonic Technology Marco Bonesi The University of Auckland, Department of Physics, New Zealand The Dodd Walls Centre for Quantum and Photonic Technology Daniel Everett The Dodd Walls Centre for Quantum and Photonic Technology The University of Auckland, Department of Chemical and Materials Engineering, New Zealand Matthew Goodwin The University of Auckland, Department of Physics, New Zealand The Dodd Walls Centre for Quantum and Photonic Technology Cushla McGoverin The University of Auckland, Department of Physics, New Zealand The Dodd Walls Centre for Quantum and Photonic Technology Sue McGlashan The University of Auckland, Department of Anatomy and Medical Imaging, New Zealand Ashvin Thambyah The University of Auckland, Department of Chemical and Materials Engineering, New Zealand Frédérique Vanholsbeeck The University of Auckland, Department of Physics, New Zealand The Dodd Walls Centre for Quantum and Photonic Technology http://arxiv.org/abs/2507.21349v1 Enhancing and Accelerating Brain MRI through Deep Learning Reconstruction Using Prior Subject-Specific Imaging 2025-07-28T21:39:36Z Magnetic resonance imaging (MRI) is a crucial medical imaging modality. However, long acquisition times remain a significant challenge, leading to increased costs, and reduced patient comfort. Recent studies have shown the potential of using deep learning models that incorporate information from prior subject-specific MRI scans to improve reconstruction quality of present scans. Integrating this prior information requires registration of the previous scan to the current image reconstruction, which can be time-consuming. We propose a novel deep-learning-based MRI reconstruction framework which consists of an initial reconstruction network, a deep registration model, and a transformer-based enhancement network. We validated our method on a longitudinal dataset of T1-weighted MRI scans with 2,808 images from 18 subjects at four acceleration factors (R5, R10, R15, R20). Quantitative metrics confirmed our approach's superiority over existing methods (p < 0.05, Wilcoxon signed-rank test). Furthermore, we analyzed the impact of our MRI reconstruction method on the downstream task of brain segmentation and observed improved accuracy and volumetric agreement with reference segmentations. Our approach also achieved a substantial reduction in total reconstruction time compared to methods that use traditional registration algorithms, making it more suitable for real-time clinical applications. The code associated with this work is publicly available at https://github.com/amirshamaei/longitudinal-mri-deep-recon. 2025-07-28T21:39:36Z Amirmohammad Shamaei Lou Alexander Stebner Lou Salome Lou Bosshart Johanna Ospel Gouri Ginde Mariana Bento Roberto Souza http://arxiv.org/abs/2507.20818v1 Comparison of Magnetic Field Characteristics among 3 Tesla MRI Scanners: An Experimental Measurement Study 2025-07-28T13:24:50Z Magnetic resonance imaging (MRI) scanners have advanced significantly, with a growing use of highfield 3 T systems. This evolution gives rise to safety concerns for healthcare personnel working in proximity to MRI equipment. While manufacturers provide theoretical Gauss line projections, these are typically derived under ideal open-environment conditions and may not reflect real-world installations. For this reason, identical MRI models can produce markedly different fringe field distributions depending on shielding and room configurations. The present study proposes an experimental methodology for the mapping of the fringe magnetic field in the vicinity of three 3 T MRI scanners. Field measurements were interpolated to generate threedimensional magnetic field maps. A comparative analysis was conducted, which revealed notable differences among the scanners. These differences serve to highlight the influence of site-specific factors on magnetic field propagation. 2025-07-28T13:24:50Z 15 pages, 15 figures presented at APSAC Conference, Salerno, Italy, June 20-23, 2025 Francesco Girardello Maria Antonietta D'Avanzo Massimo Mattozzi Victorian Michele Ferro Giuseppe Acri Valentina Hartwig http://arxiv.org/abs/2507.20793v1 Super-Resolution Radiography by Mechanical Supersampling and Model-Based Iterative Reconstruction using High-Z Photon-Counting Detectors 2025-07-28T13:03:19Z This study presents a practical and dose-efficient strategy for resolution enhancement in planar radiography, based on mechanically supersampled acquisition with high-Z photon-counting detectors (PCDs). Unlike prior event-based or cluster methods, our approach operates in true photon-counting mode and supports clinical flux rates. Using detector trajectories spanning multiple pixels and image registration-based shift estimation, we achieve sub-pixel sampling without requiring mechanical precision, while also compensating for motion and geometric instabilities. An iterative reconstruction framework based on Maximum Likelihood Expectation Maximization (MLEM) with a distance-driven ray model further enhances resolution and noise robustness. Long-range supersampling additionally mitigates pixel defects and spectral inhomogeneities inherent to high-Z detectors. Phantom studies demonstrate substantial resolution improvement and image uniformity. In comparison with a clinical mammography system, the method reveals sharper detail and more homogeneous contrast at comparable or reduced dose. The resolution gain also reduces the need for geometric magnification, enabling smaller and more cost-effective PCDs. These results establish mechanically supersampled radiography as a clinically viable approach for micron-scale imaging, with strong potential for digital mammography and other high-resolution applications and with scan times compatible with clinical workflow. 2025-07-28T13:03:19Z 8 pages, 8 figures Moritz Weigt Karl Hugo E. Helin Nicolas Karl Fix Peter Bronsert Frank Goldschmidtböing Thomas Thuering Matthias Habl Spyridon Gkoumas Thomas Stein Paul Martin Fleing Thomas Billoud Jakob Neubauer Dominik von Elverfeldt Martin Peter Pichotka http://arxiv.org/abs/2510.15875v1 A virtual airplane for fear of flying therapy 2025-07-28T02:15:06Z Fear of flying is a serious problem that affects millions of individuals. Exposure therapy for fear of flying is an effective therapy technique. However, exposure therapy is also expensive, logistically difficult to arrange, and presents significant problems of patient confidentiality and potential embarrassment. We have developed a virtual airplane for use in fear of flying therapy. Using the virtual airplane for exposure therapy is a potential solution to many of the current problems of fear of flying exposure therapy. We describe the design of the virtual airplane and present a case report on its use for fear of flying exposure therapy. 2025-07-28T02:15:06Z Proceedings of the IEEE 1996 Virtual Reality Annual International Symposium Pages 86-93 Larry F Hodges Barbara O Rothbaum Benjamin Watson G Drew Kessler Dan Opdyke 10.1109/VRAIS.1996.490515 http://arxiv.org/abs/2508.00891v1 Accelerating multiparametric quantitative MRI using self-supervised scan-specific implicit neural representation with model reinforcement 2025-07-27T03:06:49Z Purpose: To develop a self-supervised scan-specific deep learning framework for reconstructing accelerated multiparametric quantitative MRI (qMRI). Methods: We propose REFINE-MORE (REference-Free Implicit NEural representation with MOdel REinforcement), combining an implicit neural representation (INR) architecture with a model reinforcement module that incorporates MR physics constraints. The INR component enables informative learning of spatiotemporal correlations to initialize multiparametric quantitative maps, which are then further refined through an unrolled optimization scheme enforcing data consistency. To improve computational efficiency, REFINE-MORE integrates a low-rank adaptation strategy that promotes rapid model convergence. We evaluated REFINE-MORE on accelerated multiparametric quantitative magnetization transfer imaging for simultaneous estimation of free water spin-lattice relaxation, tissue macromolecular proton fraction, and magnetization exchange rate, using both phantom and in vivo brain data. Results: Under 4x and 5x accelerations on in vivo data, REFINE-MORE achieved superior reconstruction quality, demonstrating the lowest normalized root-mean-square error and highest structural similarity index compared to baseline methods and other state-of-the-art model-based and deep learning approaches. Phantom experiments further showed strong agreement with reference values, underscoring the robustness and generalizability of the proposed framework. Additionally, the model adaptation strategy improved reconstruction efficiency by approximately fivefold. Conclusion: REFINE-MORE enables accurate and efficient scan-specific multiparametric qMRI reconstruction, providing a flexible solution for high-dimensional, accelerated qMRI applications. 2025-07-27T03:06:49Z Ruimin Feng Albert Jang Xingxin He Fang Liu http://arxiv.org/abs/2507.20004v1 Solving the MIT Inverse Problem by Considering Skin and Proximity Effects in Coils 2025-07-26T16:51:49Z This paper presents an improved technique for solving the inverse problem in magnetic induction tomography (MIT) by considering skin and proximity effects in coils. MIT is a non-contact, noninvasive, and low-cost imaging modality for obtaining the distribution of conductivity inside an object. Reconstruction of low conductivity distribution by MIT requires more accurate techniques since measured signals are inherently weak and the reconstruction problem is highly nonlinear and ill-posed. Previous MIT inverse problem studies have ignored skin and proximity effects inside coils in the forward method. In this article, the improved technique incorporates these effects in the forward method. Furthermore, it employs the regularized Gauss-Newton algorithm to reconstruct the conductivity distribution. The regularization parameter is obtained by an adaptive method using the two input parameters: a coefficient and an initial conductivity distribution. The new Jacobian matrix is computed based on a standard technique. To compare the early and improved forward methods in possible medical and industrial applications with low conductivity regions, a 2D 8-coil MIT system is modeled, and image reconstruction is performed for synthetic phantoms. Results show that it is crucial to use the improved forward method for the reconstruction of the absolute conductivity values. 2025-07-26T16:51:49Z Hassan Yazdanian Reza Jafari Hamid Abrishami Moghaddam http://arxiv.org/abs/2507.19637v1 Quantifying lower-limb muscle coordination during cycling using electromyography-informed muscle synergies 2025-07-25T19:11:04Z Assessment of muscle coordination during cycling may provide insight into motor control strategies and movement efficiency. This study evaluated muscle synergies and coactivation patterns as indicators of neuromuscular coordination in lower-limb across three power levels of cycling. Twenty recreational cyclists performed a graded cycling test on a stationary bicycle ergometer. Electromyography was recorded bilaterally from seven lower-limb muscles and muscle synergies were extracted using non-negative matrix factorization. The Coactivation Index (CI), Synergy Index (SI), and Synergy Coordination Index (SCI) were calculated to assess muscle coordination patterns. Four muscle synergies were identified consistently across power levels, with changes in synergy composition and activation timing correlated with increased muscular demands. As power level increased, the CI showed reduced muscle coactivation at the knee and greater muscle coactivation at the ankle. The SI revealed a greater contribution of the synergy weights of the extensor muscles than those of the flexor muscles at the knee. In contrast, the relative EMG contribution of hip extensor and flexor muscles remained consistent with increasing power levels. The SCI increased significantly with increasing power level, suggesting a reduction in the size of the synergy space and improved neuromuscular coordination. These findings provide insight into how the central nervous system modulates its response to increasing mechanical demands. Combining synergy and coactivation indices offers a promising approach to assess motor control, inform rehabilitation, and optimize performance in cycling tasks. 2025-07-25T19:11:04Z 21 pages, 6 figures Reza Ahmadi Department of Mechanical and Manufacturing Engineering, University of Calgary, Calgary, Canada Shahram Rasoulian Human Performance Laboratory, Faculty of Kinesiology, University of Calgary, Calgary, Canada Hamidreza Heidary Department of Mechanical and Manufacturing Engineering, University of Calgary, Calgary, Canada Saied Jalal Aboodarda Human Performance Laboratory, Faculty of Kinesiology, University of Calgary, Calgary, Canada Thomas K. Uchida Department of Mechanical Engineering, University of Ottawa, Ottawa, Canada Walter Herzog Department of Mechanical and Manufacturing Engineering, University of Calgary, Calgary, Canada Human Performance Laboratory, Faculty of Kinesiology, University of Calgary, Calgary, Canada Amin Komeili Department of Mechanical and Manufacturing Engineering, University of Calgary, Calgary, Canada Human Performance Laboratory, Faculty of Kinesiology, University of Calgary, Calgary, Canada http://arxiv.org/abs/2507.19282v1 SAM2-Aug: Prior knowledge-based Augmentation for Target Volume Auto-Segmentation in Adaptive Radiation Therapy Using Segment Anything Model 2 2025-07-25T13:59:10Z Purpose: Accurate tumor segmentation is vital for adaptive radiation therapy (ART) but remains time-consuming and user-dependent. Segment Anything Model 2 (SAM2) shows promise for prompt-based segmentation but struggles with tumor accuracy. We propose prior knowledge-based augmentation strategies to enhance SAM2 for ART. Methods: Two strategies were introduced to improve SAM2: (1) using prior MR images and annotations as contextual inputs, and (2) improving prompt robustness via random bounding box expansion and mask erosion/dilation. The resulting model, SAM2-Aug, was fine-tuned and tested on the One-Seq-Liver dataset (115 MRIs from 31 liver cancer patients), and evaluated without retraining on Mix-Seq-Abdomen (88 MRIs, 28 patients) and Mix-Seq-Brain (86 MRIs, 37 patients). Results: SAM2-Aug outperformed convolutional, transformer-based, and prompt-driven models across all datasets, achieving Dice scores of 0.86(liver), 0.89(abdomen), and 0.90(brain). It demonstrated strong generalization across tumor types and imaging sequences, with improved performance in boundary-sensitive metrics. Conclusions: Incorporating prior images and enhancing prompt diversity significantly boosts segmentation accuracy and generalizability. SAM2-Aug offers a robust, efficient solution for tumor segmentation in ART. Code and models will be released at https://github.com/apple1986/SAM2-Aug. 2025-07-25T13:59:10Z 26 pages, 10 figures Guoping Xu Yan Dai Hengrui Zhao Ying Zhang Jie Deng Weiguo Lu You Zhang http://arxiv.org/abs/2505.24793v2 AFIRE: Accurate and Fast Image Reconstruction Algorithm for Geometric-inconsistent Multispectral CT 2025-07-25T09:33:17Z For nonlinear multispectral computed tomography (CT), accurate and fast image reconstruction is challenging when the scanning geometries under different X-ray energy spectra are inconsistent or mismatched. Motivated by this, we propose an Accurate and Fast Image REconstruction (AFIRE) algorithm to address such problems in the case of mildly full scan. From the continuous (resp. discrete) setting, we discover that the derivative operator (gradient) of the involved nonlinear mapping at some special points, for example, at zero, can be represented as a composition (block multiplication) of a diagonal operator (matrix) composed of X-ray transforms (projection matrices) and a very small-scale matrix. Based on these insights, the AFIRE algorithm is proposed by leveraging the simplified Newton method. Under proper conditions, we establish the convergence theory of the proposed algorithm. Furthermore, numerical experiments are also carried out to verify that the proposed algorithm can accurately and effectively reconstruct the basis images in completely geometric-inconsistent dual-energy CT with noiseless and noisy projection data. Particularly, the proposed algorithm significantly outperforms some state-of-the-art methods in terms of accuracy and efficiency. Finally, the flexibility and extensibility of the proposed algorithm are also demonstrated. 2025-05-30T16:55:23Z 39 pages, 16 figures, 1 table Yu Gao Chong Chen http://arxiv.org/abs/2507.18850v1 Estimating Sensitivity Maps for X-Nuclei Magnetic Resonance Spectroscopic Imaging 2025-07-24T23:44:02Z The purpose of this research is to estimate sensitivity maps when imaging X-nuclei that may not have a significant presence throughout the field of view. We propose to estimate the coil's sensitivities by solving a least-squares problem where each row corresponds to an individual estimate of the sensitivity for a given voxel. Multiple estimates come from the multiple bins of the spectrum with spectroscopy, multiple times with dynamic imaging, or multiple frequencies when utilizing spectral excitation. The method presented in this manuscript, called the L2 optimal method, is compared to the commonly used RefPeak method which uses the spectral bin with the highest energy to estimate the sensitivity maps. The L2 optimal method yields more accurate sensitivity maps when imaging a numerical phantom and is shown to yield a higher signal-to-noise ratio when imaging the brain, pancreas, and heart with hyperpolarized pyruvate as the contrast agent with hyperpolarized MRI. The L2 optimal method is able to better estimate the sensitivity by extracting more information from the measurements. 2025-07-24T23:44:02Z Nicholas Dwork Jeremy W. Gordon Shuyu Tang Peder E. Z. Larson http://arxiv.org/abs/2503.22139v2 Time-resolved dynamic CBCT reconstruction using prior-model-free spatiotemporal Gaussian representation (PMF-STGR) 2025-07-24T18:45:56Z Time-resolved CBCT imaging, which reconstructs a dynamic sequence of CBCTs reflecting intra-scan motion (one CBCT per x-ray projection without phase sorting or binning), is highly desired for regular and irregular motion characterization, patient setup, and motion-adapted radiotherapy. Representing patient anatomy and associated motion fields as 3D Gaussians, we developed a Gaussian representation-based framework (PMF-STGR) for fast and accurate dynamic CBCT reconstruction. PMF-STGR comprises three major components: a dense set of 3D Gaussians to reconstruct a reference-frame CBCT for the dynamic sequence; another 3D Gaussian set to capture three-level, coarse-to-fine motion-basis-components (MBCs) to model the intra-scan motion; and a CNN-based motion encoder to solve projection-specific temporal coefficients for the MBCs. Scaled by the temporal coefficients, the learned MBCs will combine into deformation vector fields to deform the reference CBCT into projection-specific, time-resolved CBCTs to capture the dynamic motion. Due to the strong representation power of 3D Gaussians, PMF-STGR can reconstruct dynamic CBCTs in a 'one-shot' training fashion from a standard 3D CBCT scan, without using any prior anatomical or motion model. We evaluated PMF-STGR using XCAT phantom simulations and real patient scans. Metrics including the image relative error, structural-similarity-index-measure, tumor center-of-mass-error, and landmark localization error were used to evaluate the accuracy of solved dynamic CBCTs and motion. PMF-STGR shows clear advantages over a state-of-the-art, INR-based approach, PMF-STINR. Compared with PMF-STINR, PMF-STGR reduces reconstruction time by 50% while reconstructing less blurred images with better motion accuracy. With improved efficiency and accuracy, PMF-STGR enhances the applicability of dynamic CBCT imaging for potential clinical translation. 2025-03-28T04:27:30Z 25 pages, 5 figures Jiacheng Xie Hua-Chieh Shao You Zhang http://arxiv.org/abs/2507.18628v1 Design and fabrication of ultrasound linear array transducer used in ultrasound endoscope 2025-07-24T17:59:16Z This report details the successful construction of an ultrasound imaging platform and the design and fabrication of a novel ultrasound endoscope probe. The projects primary objective was to establish a functional system for acquiring and processing ultrasound signals, specifically targeting minimally invasive endoscopic applications. The ultrasound imaging platform was primarily designed and developed based on Texas Instruments (TI) Evaluation Modules (EVMs). It enables the transmission of 32-channel high-voltage signals and the reception of echo signals, with on-chip signal amplification and acquisition capabilities. Furthermore, the platform integrates a complete Time Gain Control (TGC) imaging path and a ContinuousWave Doppler (CWD) path. In conjunction with host computer software, it supports imaging with linear array, convex array, and phased array probes. Concurrently, a 64-element, 5MHz center frequency, phased array linear ultrasound endoscopic probe was designed, aiming for miniaturization and optimal imaging performance. The fabrication and assembly of its matching layer, backing layer, 2-2 piezoelectric composite material, and electrodes were completed. 2025-07-24T17:59:16Z Yuan Zhang Mingtong Chen Zhengbao Yang http://arxiv.org/abs/2507.18677v1 HeartUnloadNet: A Weakly-Supervised Cycle-Consistent Graph Network for Predicting Unloaded Cardiac Geometry from Diastolic States 2025-07-24T14:31:35Z The unloaded cardiac geometry (i.e., the state of the heart devoid of luminal pressure) serves as a valuable zero-stress and zero-strain reference and is critical for personalized biomechanical modeling of cardiac function, to understand both healthy and diseased physiology and to predict the effects of cardiac interventions. However, estimating the unloaded geometry from clinical images remains a challenging task. Traditional approaches rely on inverse finite element (FE) solvers that require iterative optimization and are computationally expensive. In this work, we introduce HeartUnloadNet, a deep learning framework that predicts the unloaded left ventricular (LV) shape directly from the end diastolic (ED) mesh while explicitly incorporating biophysical priors. The network accepts a mesh of arbitrary size along with physiological parameters such as ED pressure, myocardial stiffness scale, and fiber helix orientation, and outputs the corresponding unloaded mesh. It adopts a graph attention architecture and employs a cycle-consistency strategy to enable bidirectional (loading and unloading) prediction, allowing for partial self-supervision that improves accuracy and reduces the need for large training datasets. Trained and tested on 20,700 FE simulations across diverse LV geometries and physiological conditions, HeartUnloadNet achieves sub-millimeter accuracy, with an average DSC of 0.986 and HD of 0.083 cm, while reducing inference time to just 0.02 seconds per case, over 10^5 times faster and significantly more accurate than traditional inverse FE solvers. Ablation studies confirm the effectiveness of the architecture. Notably, the cycle-consistent design enables the model to maintain a DSC of 97% even with as few as 200 training samples. This work thus presents a scalable and accurate surrogate for inverse FE solvers, supporting real-time clinical applications in the future. 2025-07-24T14:31:35Z Codes are available at https://github.com/SiyuMU/Loaded2UnNet Siyu Mu Wei Xuan Chan Choon Hwai Yap