https://arxiv.org/api/6IbEjf0yy/NKb9FwLEVVAfFJ3xI2026-06-26T06:24:40Z9008100515http://arxiv.org/abs/2505.21761v3On the Role of Demagnetizing Tensors in Arbitrary Orientations of General Ellipsoid: Implications for MRI Safety Assessment2025-09-03T09:45:41ZThis work explores the behaviour of demagnetizing tensors for general ellipsoids under arbitrary rotations in homogeneous magnetic fields. The work is motivated by the concerns in magnetic resonance imaging safety and their practical evaluation in clinical environments. Whereas demagnetizing tensor is a well-defined concept in the principal axes, its transformation under three-dimensional reorientation is often overlooked - a justifiable omission for solutions derived from Poisson equation, where the tensor can be directly rotated. However, this does not hold for common approximations, where such tensor is not explicitly defined. This work demonstrates the validity of directly rotating the orthogonal basis solutions, derived from Poissons equation, and uses the procedure to evaluate a practical approximation, based on orthogonal area-projections. The tensor rotation approach is also applied to generalize force and torque calculations for ellipsoids under three-dimensional re-orientation. The results show an exact match for translation force and torque when compared to the standard single axis rotation. Additionally, a unique connection to the well-known MRI magic angle is found as the point of convergence for prolate spheroid aspect ratios while solving the corresponding saturation field magnitudes. Finally, the evaluated approximate method was demonstrated to perform fairly across prolate and oblate spheroids. Similar approximations might extend to irregular shapes, but numerical validation would likely remain preferable due to the complexity of internal field distributions.2025-05-27T20:56:11ZThis work has been submitted to the IOP for possible publicationTomppa Pakarinenhttp://arxiv.org/abs/2509.03107v1Towards a 384-channel magnetoencephalography system based on optically pumped magnetometers2025-09-03T08:05:21ZMagnetoencephalography using optically pumped magnetometers (OPM-MEG) is gaining significant traction as a neuroimaging tool, with the potential for improved performance and practicality compared to conventional instrumentation. However, OPM-MEG has so far lagged conventional-MEG in terms of the number of independent measurements of magnetic field that can be made across the scalp (i.e. the number of channels). This is important since increasing channel count offers improvements in sensitivity, spatial resolution and coverage. Unfortunately, increasing channel count also poses significant technical and practical challenges. Here, we describe a new OPM-MEG array which exploits triaxial sensors and integrated miniaturised electronic control units to measure MEG data from up to 384 channels. We also introduce a high-speed calibration method to ensure the that the fields measured by this array are high fidelity. The system was validated using a phantom, with results showing that dipole localisation accuracy is better than 1 mm, and correlation between the measured magnetic fields and a dipole model is >0.998. We then demonstrate utility in human MEG acquisition: via measurement of visual gamma oscillations we demonstrate the improvements in sensitivity that are afforded by high channel density, and via a moviewatching paradigm we quantify improvements in spatial resolution. In sum, we show the first OPM-MEG system with a channel count larger than that of typical conventional MEG devices. This represents a significant step on the path towards OPMs becoming the sensor of choice for MEG measurement.2025-09-03T08:05:21Z31 pages, 9 figures. Submitted to Imaging NeuroscienceHolly SchofieldRyan M. HillLukas RierEwan KennettGonzalo Reina RiveroJoseph GibsonAshley TylerZoe TannerFrank WorcesterTyler HaywardJames OsborneCody DoyleVishal ShahElena BotoNiall HolmesMatthew J. Brookeshttp://arxiv.org/abs/2509.02948v1Mechanistic Insights Into How Rewiring and Bifurcation Angle Affect DK-Crush Stent Deployment2025-09-03T02:26:45ZBackground Double Kissing Crush (DKC) is a preferred two-stent technique for complex coronary bifurcation lesions. Proximal cell rewiring is routinely recommended to reduce technical failure, and DKC is considered effective across various bifurcation angles. However, it remains unclear whether this standard approach is optimal for all patients. This study investigates the interaction between bifurcation angle and rewiring configuration to identify anatomy-specific strategies. Methods Computational modeling of the DKC procedure was used to simulate 12 DKC procedures across three left main bifurcation angles (45°, 70°, and 100°) and four rewiring configurations: proximal-proximal (P-P), proximal-distal (P-D), distal-proximal (D-P), and distal-distal (D-D). Evaluation metrics included stent malapposition, side branch ostium clearance, arterial wall stress, low time-averaged endothelial shear stress, and high shear rates. Results DKC performed in wide bifurcations (100°) resulted in worse outcomes, with malapposition reaching 18%, side branch clearance down to 23%, and up to twice the exposure to adverse high shear rates compared to narrower angles. In contrast, intermediate (70°) and narrow (45°) angles generally resulted in more favorable outcomes, though optimal rewiring varied by angle. Proximal strategies, i.e. P-P and P-D, were most effective at 70°, while D-D performed best at 45°. No single strategy was consistently superior across all bifurcation angles. Conclusions DKC outcomes depend on bifurcation angle and can be optimized by tailoring rewiring strategies, challenging the current clinical understanding. These findings support anatomy-specific procedural planning and intravascular imaging to guide rewiring. This study provides a mechanistic rationale to improve clinical decision-making and tailor bifurcation interventions.2025-09-03T02:26:45Z21 pages, 4 figures in main text, 3 figures and 2 tables in supplementary materialCatheter. Cardiovasc. Interv. (2026)Andrea ColomboDario CarbonaroMingzi ZhangClaudio ChiastraMark WebsterNigel JepsonSusann Beier10.1002/ccd.70475http://arxiv.org/abs/2509.02710v1Toward a robust lesion detection model in breast DCE-MRI: adapting foundation models to high-risk women2025-09-02T18:11:18ZAccurate breast MRI lesion detection is critical for early cancer diagnosis, especially in high-risk populations. We present a classification pipeline that adapts a pretrained foundation model, the Medical Slice Transformer (MST), for breast lesion classification using dynamic contrast-enhanced MRI (DCE-MRI). Leveraging DINOv2-based self-supervised pretraining, MST generates robust per-slice feature embeddings, which are then used to train a Kolmogorov--Arnold Network (KAN) classifier. The KAN provides a flexible and interpretable alternative to conventional convolutional networks by enabling localized nonlinear transformations via adaptive B-spline activations. This enhances the model's ability to differentiate benign from malignant lesions in imbalanced and heterogeneous clinical datasets. Experimental results demonstrate that the MST+KAN pipeline outperforms the baseline MST classifier, achieving AUC = 0.80 \pm 0.02 while preserving interpretability through attention-based heatmaps. Our findings highlight the effectiveness of combining foundation model embeddings with advanced classification strategies for building robust and generalizable breast MRI analysis tools.2025-09-02T18:11:18ZGabriel A. B. do NascimentoVincent DongGuilherme J. CavalcanteAlex NguyenThaís G. do RêgoYuri MalheirosTelmo M. Silva FilhoCarla R. Zeballos TorrezJames C. GeeAnne Marie McCarthyAndrew D. A. MaidmentBruno Barufaldihttp://arxiv.org/abs/2509.01915v1The Impact of Aortic Valve Stenosis on Pulse Wave Morphology: An in silico study with 16,038 virtual subjects2025-09-02T03:21:16ZAortic valve stenosis (AVS) presents challenges in asymptomatic detection, resulting in delayed intervention. This study aims to understand how AVS affects pulse wave (PW) morphology. A PW database of 16,038 virtual subjects aged 50 to 75 was created, representing normal physiology and varying AVS degrees. All subjects were simulated using a closed-loop one-dimensional/zero-dimensional blood flow model of the entire cardiovascular system, incorporating a four-chamber heart model capable of simulating different levels of AVS by reducing the orifice area of the aortic valve. Even in cases below clinical significance, distinct PW morphology changes were observed, suggesting potential for early AVS detection using peripheral PWs from non-invasive at-home devices.2025-09-02T03:21:16ZCMBE24, Jun 2024, Arlington (Virginia), United StatesRobert D WilsonSara VardanegaJiajie ChenLucas O MüllerRachel E CloughJordi Alastrueyhttp://arxiv.org/abs/2509.01755v1A comprehensive study on the characterization of lyzed blood samples using dual-wavelength photoacoustics2025-09-01T20:16:14ZAnemia is a global health concern, prompting the need for rapid and accurate diagnostic tools, especially for vulnerable populations. Estimating the blood lysis level (LL) and oxygen saturation (SO2) are essential not only for anemia but also for other hemolytic conditions. This study explores the potential of photoacoustic (PA) spectroscopy as a quantitative tool for evaluating hemolysis in anemia diagnosis. In vitro PA measurements on human blood samples were validated through computational modeling using the discrete dipole approximation, Monte Carlo, and k-Wave acoustic simulations. The quantitative values of blood hematocrit (H), LL and SO2 have been estimated using simulated and experimental PA signals. The wavelength pairs 700-905 nm and 700-1000 nm have been found to be optimal for the simultaneous estimation of these parameters and provided H and SO2 estimations with accuracy approximately > 90% up to LL=14% and for LL = 0-30%, respectively. The correlation coefficient between the actual and evaluated lysis levels has been computed to be \approx 0.90. Further investigation is needed to enhance the robustness and clinical applicability of the developed method under an in vivo setting when both the H and LL levels are not known.2025-09-01T20:16:14ZSubhadip PaulHari Shankar PatelVatsala MisraRavi RaniAmaresh K. SahooRatan K. Sahahttp://arxiv.org/abs/2509.00435v1Quantitative imaging of 478-keV prompt gamma rays from boron neutron capture reactions2025-08-30T09:43:45ZThe accurate imaging and quantitative measurement of 478-keV prompt gamma rays are critical for advancing boron neutron capture therapy (BNCT), a promising cancer treatment. Although numerical simulations have indicated that such measurements are feasible, their practical application has proven challenging. This study introduces a gamma-ray imaging detector designed specifically for precise BNCT measurements. Using boron-rich phantom samples, we successfully imaged 478-keV gamma rays and established a linear correlation between gamma-ray production and boron concentration. Furthermore, applying this technique in a recognized BNCT treatment facility demonstrated the detector's effectiveness in monitoring boron dose distribution during neutron irradiation, both in pretreatment diagnostics and throughout the treatment process.2025-08-30T09:43:45Z35 pages, 8 figures, 2 tablesScientific Reports 15, 29310 (2025)Tetsuya MizumotoShotaro KomuraAtsushi TakadaYoshinori SakuraiToru Tanimori10.1038/s41598-025-14212-8http://arxiv.org/abs/2509.00306v1Reporte de visita médica sobre la aplicación de braquiterapia de baja tasa con Cs-137 en una paciente con cáncer de cuello uterino, estadio IIB2025-08-30T01:49:11ZDuring an academic visit to the National Oncology Institute (ION), the low-dose-rate brachytherapy procedure with Cs-137 was observed in a patient with stage IIB cervical cancer. The experience provided insight into both the clinical preparation of the patient and the dose planning process through imaging studies. Despite certain logistical limitations, strict adherence to safety protocols, disinfection procedures, and radioactive source handling was emphasized. The coordinated participation of medical physicists and technicians highlighted the practical application of physical principles, such as the use of X-rays, and underscored the relevance of an interdisciplinary approach in brachytherapy. This observation emphasized the importance of professional collaboration to ensure both patient safety and treatment effectiveness.2025-08-30T01:49:11Z4 pages, in Spanish languageMiguel Angel Mencomo Cedeñohttp://arxiv.org/abs/2509.00246v1Analysis of the Institutional Free Market in Accredited Medical Physics Residency Programs2025-08-29T21:24:16ZPurpose: The pathway to clinical medical physics involves 3 distinct stages of accredited positions, including (i) graduate education, (ii) residency training & (iii) employment. While CAMPEP accredits programs & people, the type of positions and numbers available are each more influenced by local institutional free market forces driving their choices. Methods: Understanding the overall status and trajectories of the full range of residency positions was the rationale for this survey, using publicly available data reported 2017 to 2023 timeframe. Results: Current data shows 108 therapy residency programs with 163 positions, and 38 diagnostic imaging residency programs with 39 positions. The total numbers show growth in positions by +6.5/year in therapy, and +2.1/year in imaging. There are 18 therapy and 10 imaging residencies longer than 2 years, all at university hospitals, and nearly all requiring a PhD. The number of programs that accept graduates with either MS or PhD, versus PhD only, has grown, from 69 to 105 from 2017 to 2023. There was a trend in large health networks increasing their residency numbers from 9 to 23 in this timeframe. The total 7-year growth trend was compared to numbers of graduating students, showing reasonable agreement between positions with expected inter-year variation. Conclusions: There has been growth in numbers of accredited residencies, with accompanying shifts in program types & student education expected. Most are 2-years, while there is smaller growth in longer programs with research added in, primarily at university hospitals. Major growth has occurred in residencies accepting both MS and PhD graduates, while those accepting only PhD graduates tends to be more imaging focused and at university centers. Overall, the number positions has been keeping pace with the known growth in graduates from accredited medical physics graduate programs.2025-08-29T21:24:16Z10 pages, 4 figuresAlexander NiverBrian W. Poguehttp://arxiv.org/abs/2503.04520v2Natural Frequency Dependency of Magneto-Mechanical Resonators on Magnet Distance2025-08-29T15:35:11ZThe precise derivation of physical quantities like temperature or pressure at arbitrary locations is useful in numerous contexts, e.g. medical procedures or industrial process engineering. The novel sensor technology of magneto-mechanical resonators (MMR), based on the interaction of a rotor and stator permanent magnet, allows for the combined tracking of the sensor position and orientation while simultaneously sensing an external measurand. Thereby, the quantity is coupled to the torsional oscillation frequency, e.g. by varying the magnet distance. In this paper, we analyze the (deflection angle-independent) natural frequency dependency of MMR sensors on the rotor-stator distance, and evaluate the performance of theoretical models. The three presented sensors incorporate magnets of spherical and/or cylindrical geometry and can be operated at adjustable frequencies within the range of 61.9-307.3 Hz. Our proposed method to obtain the natural frequency demonstrates notable robustness to variations in the initial deflection amplitudes and quality factors resulting in statistical errors on the mean smaller than 0.05 %. We find that the distance-frequency relationship is well described by an adapted dipole model accounting for material and manufacturing uncertainties. Their combined effect can be compensated by an adjustment of a single parameter which drives the median model deviation generally below 0.2 %. Our depicted methods and results are important for the design and calibration process of new sensor types utilizing the MMR technique.2025-03-06T15:07:54Z9 pages, 4 figures, accepted for publication in IEEE Sensors JournalJonas FaltinathFabian MohnFynn FoergerMartin MöddelTobias Knopp10.1109/JSEN.2025.3600007http://arxiv.org/abs/2504.16621v3Stable electron-irradiated [1-$^{13}$C]alanine radicals for clinically viable metabolic imaging with Dynamic Nuclear Polarization2025-08-28T21:12:08ZDissolution Dynamic Nuclear Polarisation (dDNP) increases the sensitivity of magnetic resonance experiments by $>10^4$-fold, permitting isotopically-labelled molecules to be transiently visible in MRI scans. dDNP requires a source of unpaired electrons in contact with labelled nuclei, cooled to $\sim$1K, and spin-pumped into a given state by microwaves. These electrons are usually chemical radicals, requiring removal by filtration prior to injection into humans. Alternative sources, such as UV irradiation, generate lower polarisation and require cryogenic transport. We present ultra-high-dose-rate electron irradiation as a novel alternative for generating non-persistent radicals in alanine/glycerol mixtures. These are stable for months at room temperature, quench spontaneously upon dissolution, are present in dose-dependent concentrations, and generate comparable nuclear polarisation to trityl radicals used clinically (20\%) through a novel mechanism. This process is inherently sterilising, permitting imaging of alanine metabolism \textit{in vivo}. As well as scientific novelty, this overcomes the biggest barrier to clinically translating dDNP.2025-04-23T11:23:23ZThis is the original version of a manuscript submitted to Science Advances. Per editorial policy this is not the version post the first revision, but rather the manuscript as originally submittedCatriona H. E. RooneyDepartment of Physiology, Anatomy and Genetics, University of OxfordJustin Y. C. LauGE HealthCare, Waukesha, WI, USAEsben S. S. HansenThe MR Research Centre, Aarhus University, Aarhus, DenmarkNichlas Vous ChristensenThe MR Research Centre, Aarhus University, Aarhus, DenmarkDuy A. DangThe MR Research Centre, Aarhus University, Aarhus, DenmarkKristoffer PeterssonDepartment of Oncology, University of OxfordIain D. C. TullisDepartment of Oncology, University of OxfordBorivoj VojnovicDepartment of Oncology, University of OxfordSean SmartDepartment of Oncology, University of OxfordJarrod LewisDepartment of Material Science, University of OxfordWilliam MyersCAESR, Department of Chemistry, University of OxfordZoe RichardsonDepartment of Physics, University of OxfordBrett W. C. KennedyDepartment of Physiology, Anatomy and Genetics, University of OxfordAlice M. BowenDepartment of Material Science, University of OxfordLotte Bonde BertelsenThe MR Research Centre, Aarhus University, Aarhus, DenmarkChristoffer LaustsenThe MR Research Centre, Aarhus University, Aarhus, DenmarkDamian J. TylerDepartment of Physiology, Anatomy and Genetics, University of OxfordOCMR, Cardiovascular Medicine, University of OxfordJack J. MillerThe MR Research Centre, Aarhus University, Aarhus, Denmark10.1126/sciadv.adz4334http://arxiv.org/abs/2508.05730v2A response matrix determined with a coincidence-based acquisition for correction of charge sharing spectral distortions in energy-resolved photon counting detectors2025-08-28T14:56:50ZA model-independent method is proposed to characterize and correct charge sharing spectral distortions in energy-resolved X-ray acquisitions with pixellated photon-counting detectors. The technique is based on the determination of a coincidence-based response matrix (CBRM) through a preliminary calibration with a uniform irradiation and an arbitrary polychromatic spectrum. The calibration requires the collection of the number of coincidences between a reference pixel and its neighbours for different combinations of energy bins, in order to calculate a set of charge sharing probabilities which are independent of the input spectrum. A detector response matrix is determined, which can afterwards be applied to correct other spectra acquired with the same detector and a conventional multi-comparator electronics, without introducing penalties in terms of processing time. The technique was validated with Geant4 simulations of a 1 mm thick CdTe detector and with data collected with a pixel hybrid detector made of a 300 um thick silicon sensor coupled to a Timepix4 ASIC chip. The differences between the reconstructed spectra and reference distributions of an ideal detector or a 3x3 offline clustering algorithm were evaluated in terms of mean absolute percentage errors. It is demonstrated that the response matrix can restore the spectral information with a performance close to standard clustering algorithms and is less affected by noise artifacts than analog charge summing techniques.2025-08-07T17:40:15ZVincenzo MonacoDipartimento di Fisica, Universita' degli Studi di Torino, Torino, ItalyINFN Torino, Torino, ItalyLuca BrombalDipartimento di Fisica, Universita' degli Studi di Trieste, Trieste, ItalyINFN Trieste, Trieste, ItalyPasquale DeloguINFN Torino, Torino, ItalyAlessandro FeruglioDipartimento di Fisica, Universita' di Pisa, Pisa, ItalyINFN Pisa, Pisa, ItalyMassimiliano FioriniDipartimento di Fisica e Scienze della Terra, Universita' degli Studi di Ferrara, Ferrara, ItalyINFN Ferrara, Ferrara, ItalyRenata LongoDipartimento di Fisica, Universita' degli Studi di Trieste, Trieste, ItalyINFN Trieste, Trieste, ItalyLuca MarchettiDipartimento di Fisica, Universita' degli Studi di Torino, Torino, ItalyAnna Maria PoliDipartimento di Fisica, Universita' degli Studi di Torino, Torino, ItalyLuigi RigonDipartimento di Fisica, Universita' degli Studi di Trieste, Trieste, ItalyINFN Trieste, Trieste, ItalyValeria RossoDipartimento di Fisica, Universita' di Pisa, Pisa, ItalyINFN Pisa, Pisa, Italyhttp://arxiv.org/abs/2501.07081v3Myocardial T1 mapping at 5T using multi-inversion recovery real-time spoiled GRE2025-08-28T14:49:24ZObjective: To develop an accurate myocardial T1 mapping technique at 5T using Look-Locker-based multiple inversion-recovery with the real-time spoiled gradient echo (GRE) acquisition. Approach: The proposed T1 mapping technique (mIR-rt) samples the recovery of inverted magnetization using the real-time GRE and the images captured during diastole are selected for T1 fitting. Multiple-inversion recoveries are employed to increase the sample size for accurate fitting. The T1 mapping method was validated using Bloch simulation, phantom studies, and in 16 healthy volunteers at 5T. Main results: In both simulation and phantom studies, the T1 values measured by mIR-rt closely approximate the reference T1 values, with errors less than 3%, while the conventional MOLLI sequence underestimates T1 values. The myocardial T1 values at 5T are 1553 +/- 52 ms, 1531 +/- 53 ms, and 1526 +/- 60 ms (mean +/- standard deviation) at the apex, middle, and base, respectively. The T1 values measured by MOLLI (1350 +/- 48 ms, 1349 +/- 47 ms, and 1354 +/- 45 ms at the apex, middle, and base) were significantly lower than those of mIR-rt with p<0.001 for all three layers. The mIR-rt sequence method used in our study provides high reproducibility, particularly in the middle slices, supporting its practical relevance for myocardial T1 mapping. Significance: The proposed method is feasible for myocardial T1 mapping at 5T and provides better accuracy than the conventional MOLLI sequence.2025-01-13T06:28:48ZLinqi GeYihang ZhangHuibin ZhuLang ZhangYihang ZhouHaifeng WangDong LiangHairong ZhengYanjie Zhuhttp://arxiv.org/abs/2505.23396v2Stopping cross-section for protons across different phases of water2025-08-28T14:28:36ZAccurately quantifying the energy loss rate of proton beams in liquid water is crucial for the precise application and improvement of proton therapy, whereas the slowing down of proton in water ices also plays an important role in astrophysics. However, precisely determining the electronic stopping power, particularly for the liquid phase, has been elusive so far. Experimental techniques are difficult to apply to volatile liquids, and the availability of sufficient reliable measurements has been limited to the solid and vapor phases. The accuracy of current models is typically limited to proton energies just above the energy-loss maximum, making it difficult to predict radiation effects at an energy range of special relevance. We elucidate the phase differences in proton energy loss in water in a wide energy range (0.001-10 MeV) by means of real-time time-dependent density functional theory combined with the Penn method. This non-perturbative model, more computationally-efficient than current approaches, describes the phase effects in water in excellent agreement with available experimental data, revealing clear deviations around the maximum of the stopping power curve and below. As an important outcome, our calculations reveal that proton stopping quantities of liquid water and amorphous ice are identical, in agreement with recent similar observations for low-energy electrons, pointing out to this equivalence for all charged particles. This could help to overcome the limitation in obtaining reliable experimental information for the biologically-relevant liquid water target.2025-05-29T12:33:59ZF. MatiasN. E. KovalP. de VeraR. Garcia-MolinaI. AbrilJ. M. B. ShortoH. YoriyazJ. J. N. PereiraT. F. SilvaM. H. TabacniksM. VosP. L. Grandehttp://arxiv.org/abs/2508.20678v1Start-to-end modelling of laser-plasma acceleration, beam transport and dose deposition of very high-energy electrons for radiotherapy2025-08-28T11:39:39ZRadiotherapy using very high-energy electron (VHEE) beams generated by a laser-plasma accelerator has garnered significant interest due to its dose distribution capabilities and potential to address limitations of traditional photon-based radiotherapy. To explore the feasibility of such approach, the presented study uses the parameters of OONA, the commercial <1.3\,J, <25\,fs pulse duration laser recently installed at the Weizmann Institute of Science. Through particle-in-cell simulations of laser-plasma interaction, realistic electron beams were obtained. After filtering and collimation with a beamline and arranging them into an array of pencil beams, their radiotherapeutic potential was investigated. GEANT4 simulations were used to calculate the dose deposition in water phantoms and heterogeneous phantoms with bone inserts, considering realistic beam parameters for laser-plasma accelerators. Multifield irradiation setup and the dose distribution at the isocenter through different incidence angles were studied, simulating an intensity-modulated delivery. Our findings demonstrate that polychromatic VHEE beams generated from laser-plasma accelerators, when collimated using a compact beamline of quadrupoles and dipoles, can deliver an on-axis dose with enhanced precision and uniformity. This study highlights the transformative potential of laser-plasma accelerators in advancing radiotherapy modalities, paving the way for further research and clinical implementation.2025-08-28T11:39:39Z12 pages, 8 figuresRajakrishna KalvalaAnton GolovanovArnaud CourvoisierTomer FrilingEyal KrouppLidan GrishkoVictor Malka