https://arxiv.org/api/6R+sO6kH1UUcSaneUQUH5BGHf3M 2026-09-11T00:18:22Z 24868 105 15 http://arxiv.org/abs/2608.15911v2 Identification and electrothermal modeling of persistent current switch operating regimes in a Siemens MRI magnet using external electrical measurements 2026-08-26T16:45:01Z The persistent current switch (PCS) closes the superconducting loop of an MRI magnet. During ramping it is commonly treated as a binary element: with the heater on, the PCS branch is assumed effectively open; after heater turn-off, it is assumed superconducting. Field measurements with an MPS2 service power supply (Digit Systems) on a Siemens MAGNETOM Avanto show that the PCS remains electrically active and that its resistive state can persist after heater turn-off through Joule self-heating. Dedicated maneuvers identified two resistive regimes. With the heater on, the low-voltage resistance is nearly constant; after heater turn-off, self-heating produces a power-law voltage dependence. Both are described by a single effective electrothermal relation, $R_{sw}=C\,(P_h h+V_{node}^2/R_{sw})^a$, where $h$ is the recorded heater state, spanning a 690-fold resistance range. Final-hold data from 17 Siemens magnet sessions confirmed the voltage dependence of effective PCS resistance. Since branch-current relaxation follows $τ=L/R_{sw}$, the required hold time depends on switch state rather than a fixed delay; the median margin after source settling was $3.6\,τ$. The intersection of the PCS characteristic with the external-circuit load line defines the boundary of self-sustained resistive operation. For Avanto, the predicted 0.062 V threshold lay between superconducting recovery at 0.06 V and resistive-state persistence at 0.07 V, and retrospectively classified all 24 archived segments correctly. Including both PCS resistive regimes in the dynamic model reproduced the observed current redistribution: RMSE decreased from 1.449 to 0.154 A, a factor of 9.4. This improvement required explicit representation of heater state. 2026-08-16T19:57:42Z 44 pages, 37 figures (15 main text + 22 appendix), 9 tables Volodymyr Pupkov http://arxiv.org/abs/2608.25687v1 Signal formation and induction-gap optimization in a THGEM coupled to a resistive plate anode 2026-08-26T12:05:20Z A Thick Gaseous Electron Multiplier (THGEM) detector coupled to a resistive plate anode was investigated for the first time, using an Ar:CO$_2$ (93:7) gas mixture. The resistive anode concept enabled stable operation with induction gaps significantly smaller than those typically employed in THGEM detectors, opening the possibility of improving signal formation and timing performance. The effect of the induction gap on the detector current- and charge-signal characteristics was systematically investigated. An optimal induction gap of 0.2 mm was identified based on several key signal parameters. Subsequently, detailed studies were carried out for this optimal configuration under different electric-field settings. The detector achieved a highest peak amplitude of $\sim$56 $μ$A with a rise time of $\sim$5 ns at $Δ\text{V}_{\text{THGEM}}$=1900 V, $Δ\text{V}_{\text{induction}}$=100 V, and a drift field of 1 kV/cm when irradiated with 5.9 keV X-rays. Under the same operating conditions, a time resolution of $\sim$6.0 ns was measured detecting cosmic muons. This study establishes a new THGEM detector configuration and provides guidance for its potential application in particle-detection systems, such as muon spectrometers and sampling elements of digital hadronic calorimeters. 2026-08-26T12:05:20Z Arpan Maity Luca Moleri Maryna Borysova Shikma Bressler http://arxiv.org/abs/2608.25595v1 Development and Commissioning of the Cryogenic Target Detectors for the Technical Run of the NUCLEUS Experiment 2026-08-26T10:09:06Z The NUCLEUS experiment aims to study coherent elastic neutrino-nucleus scattering (CE$ν$NS) of reactor electron antineutrinos at the Chooz nuclear power plant in France. In this work, the cryogenic target-detector module for the NUCLEUS Technical Run was developed and commissioned at the Technical University of Munich. The module comprises four gram-scale CaWO$_4$ detectors, each equipped with two Transition Edge Sensors (TES), providing a total target mass of 6.96 g. We present the design, integration, detector characterization, and X-ray-based energy calibration of these detectors. Across six characterized detectors, a mean baseline resolution of $\overlineσ_{\mathrm{BL}} = (2.79 \pm 0.60)\,\mathrm{eV}$ was achieved. The best-performing detector reached $σ_{\mathrm{BL}} = (2.16 \pm 0.02_{\mathrm{stat}})\,\mathrm{eV}$, surpassing the design goal by a factor of two and representing a state-of-the-art result for a cryogenic CaWO$_4$ detector. The target detector module was successfully operated simultaneously with the surrounding Cryogenic Outer Veto, which consists of six kg-scale Ge detectors for background discrimination, showing no measurable cross-talk. These results demonstrate the readiness of the target-detector system for the NUCLEUS Technical Run at Chooz and mark a key milestone in the development of cryogenic detectors for reactor-CE$ν$NS measurements. 2026-08-26T10:09:06Z N. Schermer the NUCLEUS collaboration H. Abele the NUCLEUS collaboration G. Angloher the NUCLEUS collaboration B. Arnold the NUCLEUS collaboration M. Atzori Corona the NUCLEUS collaboration A. Bento the NUCLEUS collaboration E. Bossio the NUCLEUS collaboration F. Buchsteiner the NUCLEUS collaboration J. Burkhart the NUCLEUS collaboration F. Cappella the NUCLEUS collaboration M. Cappelli the NUCLEUS collaboration N. Casali the NUCLEUS collaboration R. Cerulli the NUCLEUS collaboration A. Cruciani the NUCLEUS collaboration G. Del Castello the NUCLEUS collaboration S. Dorer the NUCLEUS collaboration A. Erhart the NUCLEUS collaboration M. Friedl the NUCLEUS collaboration S. Fichtinger the NUCLEUS collaboration V. M. Ghete the NUCLEUS collaboration M. Giammei the NUCLEUS collaboration J. Hakenmüller the NUCLEUS collaboration D. Hauff the NUCLEUS collaboration M. Hock the NUCLEUS collaboration F. Jeanneau the NUCLEUS collaboration E. Jericha the NUCLEUS collaboration M. Kaznacheeva the NUCLEUS collaboration H. Kluck the NUCLEUS collaboration A. Langenkämper the NUCLEUS collaboration T. Lasserre the NUCLEUS collaboration D. Lhuillier the NUCLEUS collaboration M. Mancuso the NUCLEUS collaboration B. Mauri the NUCLEUS collaboration A. Mazzolari the NUCLEUS collaboration L. McCallin the NUCLEUS collaboration H. Neyrial the NUCLEUS collaboration C. Nones the NUCLEUS collaboration L. Oberauer the NUCLEUS collaboration L. Peters the NUCLEUS collaboration F. Petricca the NUCLEUS collaboration W. Potzel the NUCLEUS collaboration F. Pröbst the NUCLEUS collaboration F. Reindl the NUCLEUS collaboration M. Romagnoni the NUCLEUS collaboration J. Rothe the NUCLEUS collaboration J. Schieck the NUCLEUS collaboration S. Schönert the NUCLEUS collaboration C. Schwertner the NUCLEUS collaboration L. Scola the NUCLEUS collaboration L. Stodolsky the NUCLEUS collaboration A. Schröder the NUCLEUS collaboration R. Strauss the NUCLEUS collaboration R. Thalmeier the NUCLEUS collaboration C. Tomei the NUCLEUS collaboration L. Valla the NUCLEUS collaboration M. Vignati the NUCLEUS collaboration M. Vivier the NUCLEUS collaboration A. Wallach the NUCLEUS collaboration P. Wasser the NUCLEUS collaboration L. Wienke the NUCLEUS collaboration http://arxiv.org/abs/2601.22727v2 A Wide Bandwidth Trans-impedance Amplifier for Picosecond-Scale SiPM Characterization in a Wide Temperature Range 2026-08-26T07:43:00Z Future high-energy physics experiments using SiPMs as photosensitive elements may require operation at low temperatures (down to 80 K) to measure single photons with high time resolution in a highly radioactive environment. This calls for a complete characterization of these sensors over a wide temperature range to find the best compromise between detector performance and cooling requirements. This paper presents the design of a transimpedance amplifier featuring high gain ($\sim 7500\ \text{V/A}$), very high speed ($<500\ \text{ps}$ rise time) and low input noise ($\lesssim 20\ \text{pA}/\sqrt{\text{Hz}}$), able to faithfully reproduce all the features of SiPM signals without affecting its noise or jitter performance. These features make the amplifier suitable for precise measurements of the time-of-arrival of single-photon signals, as well as gain and recovery time. This article provides a detailed and thorough analysis of the circuit. The network was simulated and measured in two configurations that differ in their open-loop gain and dominant pole frequencies. After selecting the best configuration for our purposes, the amplifier was characterized in detail at ambient temperature and at 80 K. Finally, we evaluated the amplifier using a SiPM operated at low over-voltage. While SiPMs are typically characterized at high over-voltage to enhance gain and minimize timing jitter, testing at low over-voltage allowed us to assess the amplifier's performance under more challenging and realistic conditions for single-photon timing. 2026-01-30T09:02:28Z Final version accepted and published by JINST 2026 JINST 21 P08026 P. Carniti C. Gotti G. Pessina D. Trotta 10.1088/1748-0221/21/08/P08026 http://arxiv.org/abs/2608.21397v2 Optical characterization and ionizing-radiation response of the BCF-20XL scintillating-wavelength-shifting fiber 2026-08-25T20:58:36Z We report optical characterization and ionizing-radiation measurements of the scintillating-wavelength-shifting (Sci-WLS) fiber BCF-20XL from Luxium Solutions. Attenuation lengths were obtained from transmitted emission spectra using a 3.0 m fiber with a spectrophotometer and described with a two-component attenuation model, yielding a long attenuation length of 6.70 m and a short attenuation length of 0.16 m. The spectral measurements also reveal wavelength-dependent attenuation, with the long component increasing with wavelength and the short component becoming less significant at longer wavelengths. The scintillation response was also characterized under radioactive alpha, beta, and gamma irradiation using SiPM-based readout, providing a benchmark measurement of the light output from BCF-20XL. 2026-08-04T16:47:01Z 12 pages, 9 figures, 2 tables; v2 clarifies wording and presentation W. Bae L. Baker K. Chen J. Cho K. Lang C. Lee E. Liang C. Mathurin C. Murthy D. Myers S. Nguyen D. Phan M. Proga M. Zalikha J. Zey http://arxiv.org/abs/2608.26202v1 Engineering of titanium transition edge sensor wafers for the BA4-90/150 receiver of BICEP Array 2026-08-25T18:03:45Z BA4-90/150, the fourth receiver to be deployed in the BICEP Array (BA) series, is a dichroic 90/150 GHz instrument targeting the frequency space where sensitivity to the CMB polarization is maximized. The receiver will be deployed in the 2026-2027 austral summer, and is set to position BA to achieve exceptionally precise measurements of cosmic microwave background (CMB) polarization and strengthen constraints on inflationary models. Recent measurements in existing BA receivers suggest that unexpectedly high loop gain in the titanium (Ti) transition edge sensors (TESs) produces excess high-frequency noise that is consequently aliased down into the science band through the time-division multiplexed readout. To reduce the loop gain, we fabricated and tested prototype Ti TES wafers containing 16 modified detector architectures designed to broaden the superconducting transition and reduce the transition steepness (alpha). We present detector performance results, which will directly inform the final integrated wafer now being designed for full receiver commissioning. 2026-08-25T18:03:45Z A. Patel P. A. R. Ade Z. Ahmed M. Amiri D. Barkats R. Basu Thakur C. A. Bischoff D. Beck J. J. Bock V. Buza B. Cantrall J. R. Cheshire J. Connors J. Cornelison M. Crumrine A. J. Cukierman E. Denison L. Duband M. A. Echter M. Eiben B. D. Elwood S. Fatigoni J. P. Filippini A. Fortes M. Gao C. Giannakopoulos N. Goeckner-Wald D. C. Goldfinger S. Gratton J. A. Grayson A. Greathouse P. K. Grimes M. Halpern S. Henderson T. D. Hoang J. Hubmayr H. Hui K. D. Irwin M. Izquierdo Poza J. H. Kang K. S. Karkare S. Kefeli J. M. Kovac C. Kuo K. Lasko K. Lau M. Lautzenhiser T. Liu S. C. Mackey N. Maher K. G. Megerian L. Minutolo L. Moncelsi Y. Nakato H. T. Nguyen R. O'Brient S. N. Paine M. A. Petroff A. R. Polish T. Prouve C. Pryke C. D. Reintsema T. Romand M. Salatino A. Schillaci B. Schmitt B. Singari A. Soliman T. St. Germaine A. Steiger B. Steinbach R. Sudiwala K. L. Thompson C. Tucker A. D. Turner C. Vergès A. G. Vieregg A. Wandui A. C. Weber J. Willmert W. L. K. Wu H. Yang C. Yu L. Zeng C. Zhang S. Zhang http://arxiv.org/abs/2608.25011v1 Aliased noise characterization and mitigation in BICEP Array 150, 220 and 270 GHz time-division multiplexed detectors 2026-08-25T18:03:44Z Early observations with the BICEP Array 150 GHz (BA2-150) and 220/270 GHz (BA3-220/270) receivers revealed detector noise equivalent temperatures (NETs) higher than expected, together with substantial detector-to-detector and module-to-module scatter. Noise measurements acquired with multiplexing off and high frequency sampling demonstrate that this excess originates from elevated high-frequency detector noise that aliases into the science band during time-division multiplexing. We show that the excess high-frequency noise is correlated with anomalously large logarithmic TES transition slopes, α, resulting in elevated electrothermal loop gain and operation near the detector stability boundary. Measurements of α indicate values substantially larger than expected, consistent with the sharper superconducting transitions introduced by the inverted TES fabrication process adopted for BA2-150 and BA3-220/270 detectors. Operational mitigation strategies were investigated through both increased multiplexing rates and elevated focal-plane operating temperatures. Faster multiplexing reduces aliasing by shifting the multiplexing Nyquist frequency beyond the excess noise roll-off, while elevated bath temperatures reduce TES electrical power and loop gain, improving detector stability and reducing NET by approximately 10%. These results demonstrate the importance of balancing TES responsivity, electrothermal stability, and multiplexed readout performance in next-generation CMB polarimeters. 2026-08-25T18:03:44Z Submitted to SPIE Astronomical Telescopes + Instrumentation 2026 S. Fatigoni P. A. R. Ade Z. Ahmed M. Amiri D. Barkats R. Basu Thakur C. A. Bischoff D. Beck J. J. Bock V. Buza B. Cantrall J. R. Cheshire J. Connors J. Cornelison M. Crumrine A. J. Cukierman E. Denison L. Duband M. A. Echter M. Eiben B. D. Elwood J. P. Filippini A. Fortes M. Gao C. Giannakopoulos N. Goeckner-Wald D. C. Goldfinger S. Gratton J. A. Grayson A. Greathouse P. K. Grimes M. Halpern S. Henderson T. D. Hoang J. Hubmayr H. Hui K. D. Irwin M. Izquierdo Poza J. H. Kang K. S. Karkare S. Kefeli J. M. Kovac C. Kuo K. Lasko K. Lau M. Lautzenhiser T. Liu S. C. Mackey N. Maher K. G. Megerian L. Minutolo L. Moncelsi Y. Nakato H. T. Nguyen R. O'Brient S. N. Paine A. Patel M. A. Petroff A. R. Polish T. Prouve C. Pryke C. D. Reintsema T. Romand M. Salatino A. Schillaci B. Schmitt B. Singari A. Soliman T. St. Germaine A. Steiger B. Steinbach R. Sudiwala K. L. Thompson C. Tucker A. D. Turner C. Vergès A. G. Vieregg A. Wandui A. C. Weber J. Willmert W. L. K. Wu H. Yang C. Yu L. Zeng C. Zhang S. Zhang http://arxiv.org/abs/2608.24806v1 Effects of Quantum Noise and Source Blurring on Dark-Field Signal Retrieval in X-ray Speckle-Based Imaging 2026-08-25T16:49:35Z X-ray speckle-based dark-field imaging offers high sensitivity to sub-pixel structural features, yet its quantitative reliability in clinical and preclinical settings remains constrained by low photon flux and finite focal spot sizes. However, how hardware-induced noise and source blurring propagate through retrieval algorithms to degrade signal integrity is not fully understood. Here, we systematically evaluate algorithm robustness quantified by signal linearity, sensitivity, and bias under photon starvation and source blurring across two mathematically distinct frameworks: differential-based intrinsic tracking (Low-Coherence System, LCS) and patch-wise explicit tracking (X-ray Speckle-Tracking Speckle-Vector-Tracking, XST-XSVT). Our experimental results demonstrate that input speckle pattern distortions propagate through retrieval algorithms in fundamentally different ways depending on algorithm architecture. As an example, using our setup, under severe photon starvation, derivative noise amplification in LCS causes its dark-field signal linearity and sensitivity to drop precipitously, while sharply elevating baseline bias. In contrast, XST-XSVT restricts these losses for sensitivity while maintaining a stable baseline, as its patch-wise variance calculation inherently suppresses stochastic noise. Similarly, under blur-limited conditions , source blurring washes out the speckle pattern, directly reducing dark-field sensitivity for both LCS and XST-XSVT. This characterization establishes operational boundaries for low-power and low-coherence X-ray systems, guiding algorithm selection and framework optimization to realize quantitative dark-field imaging in preclinical and clinical applications. 2026-08-25T16:49:35Z 18 pages 9 figures Hunwoo Lee Jingcheng Yuan Mini Das http://arxiv.org/abs/2608.26198v1 Agentic AI for operating scientific instruments for nanoscale characterization 2026-08-25T14:55:40Z Operating a scientific instrument such as an atomic force microscope (AFM) requires continuous expert decision-making. A trained user defines the experimental intent, translates it into instrument commands, assesses incoming data, adjusts imaging parameters, and post-processes the final image. Existing automation usually addresses only parts of this workflow through hard-coded routines, task-specific controllers, or trained machine-learning models. Here we present an agentic-AI framework that operates the executable part of the AFM workflow using a general-purpose, tool-augmented large language model connected to instrument functions through the Model Context Protocol (MCP). The framework consists of 3 MCP-based agents: AFM Messenger converts natural-language instructions into checked instrument commands; AFM Pilot assesses image quality through a large language model (LLM) and, if necessary, adapts imaging parameters; and AFM Doctor diagnoses image artifacts and applies transparent post-processing from a pre-approved tool set. Because the language model performs image assessment rather than a fixed scalar objective or external optimizer, the same strategy can be applied across sample types and imaging modes without specific retraining. Safe hardware operation is enforced through an ambiguity check layer before execution. Benchmarking against fine-tuned and off-the-shelf tool-using models shows that this guarded execution layer, rather than model capability alone, reduces wrong-command execution to zero. In live experiments on different samples, AFM Pilot matched expert operators in image quality, iteration count, and tuning time, with no significant difference. These results demonstrate a safe route to agentic operation of scientific instruments, where experimental intent remains human-defined while command execution, image-based tuning, and post-processing are delegated to AI agents. 2026-08-25T14:55:40Z Zahra Ayar Marcos Penedo Mahdi Mehdikhani Nahid Hosseini Prabhu Prasad Swain Georg E. Fantner http://arxiv.org/abs/2608.21581v2 Reconstructing High-Fidelity Light Yield Maps for Surface LArTPCs Using Crossing Cosmic Muons 2026-08-25T14:16:21Z Photon detection systems in liquid argon time projection chambers provide prompt scintillation light information that can improve triggering, timing, calorimetry, and interaction reconstruction. These applications require an understanding of the spatial dependence of the detected light yield (LY), which is not adequately described by a single detector-wide average. We present a method for reconstructing voxelized 3D light yield maps in surface LArTPCs using crossing cosmic muons. For each selected muon, the path length through every crossed voxel is converted to deposited energy using a minimum ionizing particle approximation, producing a linear system relating the unknown voxel light yields to the total detected photon signal. The resulting inverse problem is solved using nonnegative least squares, with an additional $L_2$ smoothness penalty used to stabilize weakly constrained voxel values. The method is studied in simulation using the ProtoDUNE-VD detector geometry and photon detection system. Reconstructed maps recover the dominant spatial structure of a visibility-based truth reference and respond as expected to changes in the photon detector configuration. Smoothness regularization reduces zero-valued voxels and localized fluctuations while leaving the detector-average light yield approximately unchanged at the selected regularization strengths. The voxel-wise RMSE relative to the truth reference is reduced by approximately 44% for the regularized reconstruction. 2026-08-21T19:32:17Z 24 pages, 15 figures Alex Heindel Wei Shi Angelo Ralaikoto Shuaixiang Zhang Laura Paulucci Franciole Marinho Maressa Pimenta Sampaio Sanskar Jain http://arxiv.org/abs/2608.24564v1 Fabrication and validation of a MEMS Electron Monochromator 2026-08-25T13:46:18Z Electron monochromators are essential components for many applications in electron microscopy but are often large and complex to operate instruments. In prior work, we presented a concept for monochromation using the fringe fields around miniature electrostatic elements. This concept could be realized with a compact monochromator design that requires tuning of only 7 power supplies. In this manuscript, we present the fabrication using MEMS technology and experimental testing of such a miniaturized monochromator. The fabrication scheme involves Bosch deep reactive ion etching to shape lens and deflector electrodes and focused ion beam milling to create nanoscale apertures in a silicon nitride membrane. All elements are then coated, precisely aligned and, after stacking, wire bonded to connect to external power supplies. We use the assembled prototype to measure the electron-beam energy spread in a scanning electron microscope demonstrating the energy filtering capability of the device. These results highlight the possibilities for making miniaturized electron optics instruments using alignment and stacking of different MEMS fabricated elements, representing an important step towards the development of compact, lightweight, high performance electron-beam instrumentation. 2026-08-25T13:46:18Z M. J. Adriaans J. P. Hoogenboom A. Mohammadi-Gheidari http://arxiv.org/abs/2608.24553v1 Calibration of the discriminator threshold of the Amplifier-Shaper-Discriminator Chip for the Upgrade of the ATLAS Muon Drift-Tube Chambers for the High-Luminosity LHC 2026-08-25T13:37:56Z The read-out electronics of the ATLAS muon drift-tube chambers will be upgraded for operation at the High-Luminosity Large Hadron Collider. Part of the upgrade is the introduction of new amplifier-shaper-discriminator chips. This note presents the calibration of the discriminator threshold of the new chip in terms of primary electron signals using data from small-diameter muon drift-tube chambers operated in a monoenergetic, high-energy muon beam. The discriminator threshold intended for use in the ATLAS muon spectrometer corresponds to the signal of 20 primary electrons. 2026-08-25T13:37:56Z Oliver Kortner Nick Meier Julia Okfen http://arxiv.org/abs/2505.21902v2 Rise Time and Charge Collection Efficiency of Graphene-Optimized 4H-SiC p-i-n Detector 2026-08-25T11:20:40Z Silicon carbide detectors exhibit good detection performance and are being considered for detection applications. However, the presence of surface electrode of detector limits the application of low-penetration particle detectors, photodetectors and heavy-ion detection. A graphene-optimized 4H-SiC detector has been fabricated to expand the application of SiC detectors.Its electrical properties and the charge collection performance of α particles are reported. The effective doping concentration of lightly doped 4H-SiC epitaxial layer is about 4.5\times10^{13}cm^{-3}, approaching the limit of the lowest doping level by the SiC epitaxial growth technique. The rise time of the graphene-optimized ring electrode detector is reduced by 24% at 200 V, compared to ring electrode detector. The charge collection efficiency (CCE) of graphene-optimized 4H-SiC PIN is 99.22%. When the irradiation dose is 2\times10^{11} n_{eq}/cm^2, the irradiation has no significant impact on the rise time and uniformity of the rise time for the graphene-optimized 4H-SiC detectors. This study proves that graphene has a certain radiation resistance. Graphene-optimized 4H-SiC detectors can not only reduce the signal rise time, but also improve uniformity of signal rise time and stability of charge collection. This research will expand the application of graphene-based 4H-SiC detectors in fields such as low energy ions, X-ray, UV light detection, particle physics, medical dosimetry and heavy-ion detection. 2025-05-28T02:31:58Z Zhenyu Jiang Xuemei Lu Congcong Wang Yingjie Huang Xiaoshen Kang Suyu Xiao Xiyuan Zhang Xin Shi http://arxiv.org/abs/2507.17068v2 Fast 4D-STEM-based phase mapping for amorphous and mixed materials 2026-08-25T07:39:26Z Interpretation and mapping strategies for 4D-scanning transmission electron microscopy (4D-STEM) are well-developed for crystalline materials, yet in the case of amorphous and mixed materials it is significantly more challenging to separate different phases. Non-negative matrix factorization (NMF) in principle would allow separation of 4D-STEM data into components with interpretable diffraction signatures and intensity maps, independent of the crystalline, amorphous or mixed nature of the material. However, adoption of NMF in this field is hampered by large datasets and conceptual hurdles: NMF tackles a non-convex optimization problem, requiring iterative algorithms. Additionally, the stopping condition has to be chosen carefully. In this work, we show that the factorization of large 4D-STEM datasets can be drastically accelerated using a QB decomposition (i.e. randomized NMF or RNMF), leading to much shorter time per iteration. This allows structure-independent phase mapping on very large 4D-STEM datasets. We validate this approach on a synthetic literature dataset (mixed ZrCuAl), before mapping a thin TiO$_2$ layer on top of SiO$_2$, and an interface between a lithium-ion cathode and solid state electrolyte. We also demonstrate that, before using NMF to transform the data on an interpretable, nonnegative basis, principal component analysis (PCA) can be used for fast exploratory analysis to assess dataset dimensionality and linearity. 2025-07-22T23:07:48Z Andreas Werbrouck Nikhila C. Paranamana Xiaoqing He Matthias J. Young 10.1093/mam/ozag079 http://arxiv.org/abs/2608.23931v1 Photogrammetry for Precise, Rapid Module Alignment in the Mu2e Tracker 2026-08-25T00:42:32Z We discuss a photogrammetry technique to measure the inter-module alignment in the Mu2e tracker. The tracker consists of 216 modules each with 96 straw tubes. The relative alignment of the straw tubes in a module was previously measured by an X-ray technique, which tied the straw tube measurements to three fiducials per module. The photogrammetry technique uses an array of 15 cameras to make precise position measurements of the module fiducials in a global tracker coordinate system. Measurements of the fiducial position and radius on the camera's CCD yield ~$25 μm$ resolution in the transverse directions and < $300 μm$ along the camera axis. In addition, we describe a procedure to combine the images with "local" mechanical measurements to yield $<100 μm$ precision along the camera axis. The technique offers a touch-less, rapid (~1 hour), and affordable metrology approach within the requirements of the tracker. 2026-08-25T00:42:32Z 15 pages, 15 figures. Submitted to Nuclear Instruments and Methods in Physics Research A Dylan Palo Mamta Jangra Aseet Mukherjee Klara Northrup